Operating Manual 

www.geotechuk.com

 

LFWATOM01  Iss.12 

© QED Environmental Systems Ltd. 

 

A TE X

 

D I PM E TE R

 

Operating notes 

To power on the dipmeter, press and release the ON/OFF button, the meter will beep briefly. It will remember 
the last sensitivity setting when switched on. 

To power off the meter, press and hold the ON/OFF button for 3 seconds. It will power off automatically if not 
used within 5 minutes. 

Sensitivity  

If you are using the dipmeter in pure water, you may have to increase the sensitivity, to do this, press the + 
button. 

If you are using the dipmeter in cascading or saline water, you may have to decrease the sensitivity, to do this, 
press the – button. 

Reel Lock 

To lock the reel, turn the plastic knob on the rear clockwise until locked. 

Equipment Check 

 

Test circuit and battery by pressing the ON button.  Make sure the panel securing knobs are tight. If the 
unit does not sound, replace the battery (one 9 volt Duracell battery MN1604) in the drawer on the 
faceplate and repeat. 

 

Test the tape and probe by shorting out the centre conductor and probe body on the stud on the back 
axle of the unit. The buzzer and light should activate; if not, adjust the sensitivity and repeat. Make sure 
the unit is on. 

 

Test the unit in tap water before going out to the field. DO NOT use distilled or de-ionised water. 

Use in the Field 

Note: Ensure that the panel securing nuts are tight before use. 

 

Reel the tape down the well carefully, avoiding the edge of the casing. Hang the unit on the casing, where 
possible, and run the tape over the frame leg to avoid cuts and nicks. 

 

When the unit sounds, carefully measure the depth to water from your reference point by slowly 
lowering and raising the probe to the air/water interface. Raise the probe, shake off the water and repeat 
the measurement. In wells with cascading water, reduce the sensitivity by pressing the (-) button to avoid 
false signals. 

 

The probe is rated to full depth and can be used to measure depth to the bottom of the well. Reel the 
tape until the probe touches the bottom and the tape becomes slack.  DO NOT let the probe fall under 
gravity or it will be damaged when it hits the bottom of the well. DO NOT use the unit to measure sand 
backfill as the tape and probe may get "locked" in the backfill. 

 

Wind the tape back onto the reel, removing any excess moisture and dirt. 

Note: Intermittent beep indicates unit ON.  Beeping shuts off automatically 5 minutes after last use. 

Summary of Contents for Atex Dipmeter

Page 1: ...ATEXDIPMETER OPERATING MANUAL...

Page 2: ...und when the unit is tested 7 Continuous Sound when the unit is turned on or probe removed from water 7 To test the system 7 General 8 EU Declaration of Conformity 9 Appendices safety instructions 10...

Page 3: ...ow to troubleshoot the meter in case you have any issues and includes the declaration of conformity Hazard warnings and safety symbols Information in this manual that may affect the safety of users an...

Page 4: ...e of practice IEC 60079 19 If the equipment is likely to come into contact with aggressive substances then it is the responsibility of the user to take suitable precautions that prevent it from being...

Page 5: ...NT REAR FRONT FASCIA A Battery compartment H Date time stamp of manufacture B Probe I On Off switch C Instrument test function J Increase sensitivity button D Maximum depth K Decrease sensitivity butt...

Page 6: ...he buzzer and light should activate if not adjust the sensitivity and repeat Make sure the unit is on Test the unit in tap water before going out to the field DO NOT use distilled or de ionised water...

Page 7: ...yl and 1 3 denatured alcohols freons TF TE Rinse well with water and let it dry before putting the panel back in Do NOT use abrasives partially halogenated hydrocarbons or ketones to clean the reel If...

Page 8: ...ATEX DIPMETER General Avoid sharp edged casing Avoid entanglement with other equipment in boreholes and wells Do not use as guide to backfilling with sand etc Instrument may get locked in sand Rewind...

Page 9: ...Page 9 of 18 Operating Manual www geotechuk com LFWATOM01 Iss 12 QED Environmental Systems Ltd ATEX DIPMETER EU DECLARATION OF CONFORMITY...

Page 10: ...e est certifi comme tant adapt l installation en zone dangereuse II 2 G Ex ia IIB T4 Gb Cet appareil est certifi pour utilisation temp rature ambiante comprise entre 20 et 40 C mais la sonde peut tre...

Page 11: ...infiltration d eau Test du syst me Maintenir la goupille centrale de la sonde contre le goujon au dos de l axe de bobine et en m me temps toucher la vis du cadre l aide de la sonde Si le syst me fonct...

Page 12: ...dass das Ger t in Kontakt mit aggressiven Substanzen kommt liegt es in der Verantwortung des Benutzers geeignete Vorsichtsma nahmen zu treffen die negative Auswirkungen auf das Ger t verhindern und s...

Page 13: ...geschaltet oder Sonde aus dem Wasser entfernt wird berpr fen dass die Spitze des Sondenleiters sauber ist Auf berm ige Feuchtigkeit an der R ckseite des Elektronikfelds pr fen Sonden Bandverbindung un...

Page 14: ...zate non vengano danneggiate Il misuratore di profondit stato certificato in base alla classificazione delle aree pericolose II 2 G Ex ia IIB T4 Gb Questa apparecchiatura certificata unicamente per l...

Page 15: ...otture e perdite che potrebbero provocare l ingresso di acqua Per povare il sistema Tenere il perno centrale della sonda contro il montante sul retro dell asse della bobina e toccare contemporaneament...

Page 16: ...garantizar que no se comprometa el tipo de protecci n El dipmeter ha sido homologado en la clasificaci n de zonas peligrosas como II 2 G Ex ia IIB T4 Gb El equipo s lo est homologado para utilizarse a...

Page 17: ...rar agua Para probar el sistema Sostenga el borne central de la sonda contra el esp rrago de la parte trasera del eje del carrete y a la vez toque el tornillo en el bastidor con el cuerpo de la sonda...

Page 18: ...ectronic Equipment directive WEEE makes producers responsible from July 1st 2007 in meeting their obligations with the fundamental aim of reducing the environmental impact of electrical and electronic...

Reviews: