8
VERIFICARE PRECIZIE
ACCURACY CHECK
Set up instrument in the middle of two walls
which are about 5 m apart. Press buttons H and
V1. Mark a visible laser cross on one wall. Turn
unit to opposite wall and mark a laser cross.
Repeat measurements with distance of about
0.6 m to one wall and about 4.4 m to second
wall. Deviation between two measurements
taken from the centre and two measurements
taken at 0.6 m and 4.4 m must not exceed
3 mm.
Testing accuracy of horizontal line (end to
end)
Set up instrument about 5 m from a wall. Press
buttons H and V1. Mark laser cross on the wall.
Turn instrument until the laser cross has moved
about 2.5 m to the left side and check if hori-
zontal line is within ± 2 mm of laser cross
marked on wall. Repeat measurement by tur-
ning instrument to the right side
Testing accuracy of vertical line (end to end)
Set up instrument about 5 m from a wall. Fix
a plumb line of 2.5 m length to the wall, using
a plumb bob. Press button V1 and bring laser
line into coincidence with the plumb line. Devi-
ation between laser line and plummet cord from
top to bottom must not exceed ± 1.5 mm.
Configura
þ
i instrumentul în mijlocul dintre dou
ã
ziduri care sunt de aproximativ 5 m distanta .
Ap
ã
sa
þ
i butoanele H si V1. Marcati o cruce laser
vizibil pe un perete. Roti
þ
i unitate de la peretele
opus si marcati o cruce laser.
Se repet
ã
m
ã
sur
ã
torile cu distan
þ
ã
de aproximativ
0,6 m de la un perete si aproximativ 4,4 m de al
doilea perete. Devia
þ
ia dintre doua masuratori
luate de la centru
þ
i dou
ã
m
ã
sur
ã
tori luate la 0,6
m si 4,4 m nu trebuie s
ã
dep
ã
þ
easc
ã
3 mm.
Precizie Testarea linie orizontal
ã
(de la la capat la
capat)
Configura
þ
i instrumentul la circa 5 m de un
perete. Ap
ã
sa
þ
i butoanele H si V1. Marcati laser
crucea pe perete. Roti
þ
i instrumentul pân
ã
când
cruce laser s-a mutat aproximativ 2,5 m, spre
partea stâng
ã
þ
i verifica
þ
i dac
ã
orizontal
ã
Linia
este în limitele de ± 2 mm dintre crucea laser
marcat
ã
pe perete. Repetati m
ã
surarea prin
rotirea instrumentului spre partea dreapta
Precizie Testarea linie vertical
ã
(un cap
ã
t la altul)
Configura
þ
i instrumentul la circa 5 m de un
perete. Fixati o linie verticala la 2,5 m lungime de
perete, folosind un fir cu Plumb. Ap
ã
sa
þ
i butonul
V1
þ
i aduceti laser de linie în coinciden
þ
ã
cu firul
cu plumb. Abaterea între linia de centrare laser si
cablul dintre partea de sus si jos, nu trebuie s
ã
dep
ã
þ
easc
ã
± 1,5 mm.