52
•
Operational Reference
Genie Nano-CL Series™
•
TriggerOverlap=Off and ExposureMode=TriggerWidth
Diagram Conditions:
•
TriggerMode=On
•
ExposureMode=TriggerWidth
•
TriggerActivation=RisingEdge
•
TriggerDelay=0
•
TriggerSelector=FrameStart
•
ExposureAlignment=Synchronous
Trigger Input
Exposure 1
Readout 1
Frame 1 Active period
Exposure 2
Readout 2
Frame 2 Active period
Frame Exposure
Frame Readout
TriggerOverlap= Off
and
ExposureMode=TriggerWidth
Exclusion Region