BCB US Rev.00 01-2019
pag. 35
8.2.5
– CLEANING THE CORE PROBE
Carry out this operation at every cycle.
APPLIANCE STATUS:
-
I/O main switch
in position “O” (OFF);
AUTHORIZED PERSONNEL
Appliance operator.
METHOD
The core probe (ref. 1 fig. 10) must always be cleaned before a new cycle so as to avoid polluting the product in any way.
Remove all residue by means of a sponge soaked in neutral detergent. Rinse with lots of water and apply a sterilising
product.
Fig. 10
Do not pull the probe wire; you may damage it.
The probe has a very pointed end. During cleaning always wear safety gloves and take the utmost
care.
1