Insert the inner tube (24) into the Instrument Holder; insert the lever into the slot on
the closed side. Before doing that, make sure that the spiral spring (23) is placed on
its right position inside the Instrument holder.
Insert the scope holder (red plastic ball with inlay) into the mount of the inner tube
(24). Place the
ball spanner with the pins into the slot. Push it to the end, and turn the
spanner 90° until the pins are adapted to the slots.
III-3 Connecting the Instrument Holder with the pneumatic central stand
IMPORTANT: The system must be without pneumatic pressure
The Instrument Holder is prepared according to part III-2.
The pneumatic central stands housing ends in a lock nut distally (16). This nut is
marked with a dot-point, the housing is marked with an arrow. Turn the nut counter-
clockwise until both are in one line. Also the Instrument holder is marked with a dot-
point. If this dot-point is in one line with the dot-point of the nut and the arrow, the
Instrument holder can be inserted into the housing. Fix the Instrument Holder by
turning the nut clockwise 45°.