The surgeon select the lever (17) position, “down or on top”, before he place the
endoscope in the tightening ball. The positioning is to realize by a half turn.
IMPORTANT!
For laparoscopic procedures:
Before fixing the endoscope with the Pneumatic Endoscope Holder all
trocars should be on place, otherwise this activity can destroys the
endoscope.
V - 5 On the end of the procedure
• Stop the pressure supply. Close the swivel valve on the pressure cylinder or
disconnect the wall connector. Press the lever (17).
• Open the tightening ball fixation and remove the endoscope.
• Disconnect the instrument holder from the pneumatic central stand.
• Disassemble the instrument holder and the tightening ball, clean it and prepare
it for sterilization. The pneumatic central stand then to be processes according
to chapter IV.
• Suitable preparation according to chapter IV.