6.5 - Analogue output setup
−AnlgOP−
6.1C
Low scale value for analogue O/P
B
Data source for analogue O/P
6.5A
High scale value for analogue O/P
C
Cal low analogue O/P
E
Yes
No
Calibrate analogue O/P?
D
Cal high analogue O/P
F
A
_ _ Data source for analogue O/P
scrolls across the bottom row, and the cur-
rent analogue output data source appears in the top row. Use the
and
buttons to select an option from:
Sum
,
Input 1
, or
Input 2
, and then press
.
See Appendix A for more information on Summing.
B
_ _ Low scale value for analogue O/P
scrolls across the bottom row, and the
currently selected low scale display value appears in the top row. Adjust this
value as required using the
and
buttons, and then press
.
C
_ _ High scale value for analogue O/P
scrolls across the bottom row, and the
currently selected high scale display value appears in the top row. Adjust this
value as required using the
and
buttons, and then press
.
Factory analogue output calibration is precisely set before shipping this instrument.
Do not recalibrate your analogue output unless instructed by the manufacturer.
D
_ _ Calibrate analogue O/P?
scrolls across the bottom row, and
No
appears in
the top row. Use the
and
buttons to select
No
or
Yes
, and then press
.
If you selected
Yes
, connect a mA or volt meter across the analogue out-
put connector (see section 4), and then continue to 6.5E.
If you selected
No
, skip to 6.5G now.