Introduction
Vscan Extend
–
User Manual
8-5
5721203-100
Rev. D
Audit logging content
The following events are captured by the audit logging of the
Vscan Extend:
1. Device
Start/Shut-down
time
2. Incorrect
password attempts
3. Addition and Modifications to system configuration,
including:
a. DICOM
connection
b. Windows
Share
connection
c. Application software installation/uninstallation/software
upgrade
d. Registration of partnering apps
4. Events related to Patient Data, including details of:
e. Data access Date/Time including type of action
(addition, deletion, modification, reviewed) excluding
the PC connect scenario
f.
Details of applications used (Bladder app, Lung
protocol, etc.)
5. Data
Export
a. Data export Date/Time including end point and Wi-Fi
SSID information
b. Data
Backup/Restore
detail
6. Invalid Device PIN attempt
7. Kiosk admin mode login success and failure
Management of Audit logs
Back up of audit logs
Back up of audit logs can be done by exporting audit logs to an
external device. The export happens through USB export.
Remember that audit logs and exported audit log files do contain
Personal Information (PI) and must be handled according to
applicable regulations and guidelines for handling of PI/PHI.
Note that exported audit log files are stored unencrypted.
Summary of Contents for Vscan Extend
Page 4: ...i 2 Vscan Extend User Manual 5721203 100 Rev D This page intentionally left blank ...
Page 56: ...Safety 2 30 Vscan Extend User Manual 5721203 100 Rev D ...
Page 94: ...Preparing Vscan Extend for Use 3 38 Vscan Extend User Manual 5721203 100 Rev D ...
Page 212: ...Vscan Extend Maintenance 6 18 Vscan Extend User Manual 5721203 100 Rev D ...
Page 226: ...Appendix 7 14 Vscan Extend User Manual 5721203 100 Rev D ...
Page 248: ...Privacy and Security 8 22 Vscan Extend User Manual 5721203 100 Rev D ...
Page 251: ......