Vscan Extend Maintenance
6-2
Vscan Extend
–
User Manual
5721203-100
Rev. D
System care and maintenance
Overview
The Vscan Extend requires regular care and maintenance to
function safely and properly.
The expected service life of the Vscan Extend is 5 years.
To ensure that the Vscan Extend constantly operates at
maximum efficiency, we recommend that the following
procedures be observed as part of the customer’s internal
routine maintenance program.
CAUTION
The user must ensure that safety inspections are performed at
least every 12 months according to the requirements of the
patient safety standard EN/ES/IEC 60601-1 and its national
deviations or according to the requirements of the IEC62353
standard, medical electrical equipment – recurrent test and test
after repair of medical electrical equipment.
Only trained persons should perform the safety inspections.