Section 10-3 - Why Perform Maintenance Procedures?
P R E L I M I N A R Y
10-6-8
Isolated Patient Lead (Source) Leakage – Lead-to-Lead
When using any calibrated Electrical Safety Analyzer tool compliant with AAMI/ESI 1993 or IEC 60601
or AS/NZS 3551, switch the meter's function selector to the LEAD-LEAD position (or equivalent function
in the device).
Select and test each of the ECG lead positions (except ALL) on the LEAD selector, testing each to the
power and ground condition combinations found in:
•
or
•
as applicable.
Record the highest leakage current measured.
10-6-8-1
Lead-to-Lead Leakage Test Record
below shows a typical format for recording the patient lead-to-lead leakage current.
Measurements should be recorded from each lead combination under each set of test conditions
specified in:
•
or
•
as applicable.
Record all data on the EQC inspection certificate; also known as Patient Auxiliary Current.
1) Connect Safety analyzer to wall AC power outlet.
2) Plug the equipment under the test power cable into the receptacle on the meter's panel.
3) Connect the ECG cable to the Vivid™ S60/Vivid™ S70 ultrasound system and the Patient leads to
the analyzer.
4) Select the Patient lead leakage function on the meter.
5) Test opening and closing the ground with the Vivid™ S60/Vivid™ S70 ultrasound system ON and
OFF, as indicated in
on page 10-17, as applicable.
NOTE:
Refer to the safety analyzer’s user manual that will be used to perform the tests.
Table 10-16 Typical Data Format for Recording Patient Lead-to-Lead Leakage
Unit under test____________________________________ Date of test:_____________
Test Conditions
Patient Lead or Combination Measured
System Power
Grounding/PE
RA-LA
LA-LL
LL-RA
System Off
open
System On
(Transmit)
open
Summary of Contents for Vivid S60N
Page 4: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL ii ...
Page 5: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL iii ...
Page 6: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL iv ...
Page 7: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL v ...
Page 8: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL vi ...
Page 9: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL vii ...
Page 10: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL viii ...
Page 11: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL ix ...
Page 12: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL x ...
Page 13: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL xi ...
Page 14: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL xii ...
Page 477: ......