![GE Vivid 3 Service Manual Download Page 302](http://html.mh-extra.com/html/ge/vivid-3/vivid-3_service-manual_128704302.webp)
GE M
EDICAL
S
YSTEMS
D
IRECTION
2300164-100, R
EVISION
7
V
IVID
™ 3 P
RO
/V
IVID
™ 3 S
ERVICE
M
ANUAL
7-58
Section 7-3 - Performing Front End (FE) Diagnostics
7-3-9-6
Advanced Board Tests
The advanced board tests check two additional parts of the MUX Board:
TEE probe control
and
Monitoring
, as follows:
-
TEE probe control enables you to check different positions of the probe; by changing the
probe angle, you are able to verify its normal operation.
-
Monitoring enabled you to check the HVPS and LVPS voltages, and the AC Power Input, as
well as temperatures.
7-3-9-6-1 TEE Probe Control
The TEE probe control test checks the position setting of the probe, as well as its normal operation.
1) Access the MUX tab of the FB Diagnostics dialog box, as described in the
Accessing the MUX
Diagnostic Test Options
section, on page 7-47.
2) Trackball to the
Advanced board tests
button and press
Select
. The
Advanced System Tests
options are displayed, as shown in
Figure 7-34
on page 7-58.
Figure 7-34 MUX Advanced System Tests Options
3) Select the radio button next to the appropriate socket.
4) Trackball to the
TEE Probe Control
button and press
Select
. The
TEE Probe Test
dialog box is
displayed, as shown in
Figure 7-35
, with all the probe settings.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com