Section 7-3 - Performing Front End (FE) Diagnostics
Figure 7-18 Report Dialog Box - FLA Jumper Test (Closed)
The
Report
dialog box contains the following information:
•
Test Result Message:
Displays the test result.
•
FB Elements:
Displays a graphic representation of the front board elements, as follows:
•
Each element is represented by a button (highlighted in green) that displays the element,
and the channel number that is associated with that element.
•
Each button is highlighted yellow when selected.
•
Each button is highlighted red if errors were detected for one or more elements.
•
Checkboxes:
Enable you to manipulate bad channels:
•
Show Only Bad Channels:
Select this checkbox to display only bad channels in the
Display
area.
•
Mark bad channel:
Select this checkbox to manually mark the currently selected channel
as bad.