Chapter 7 - Diagnostics/Troubleshooting
7-45
7-3-8-8
RAM Test
The RAM test enables you to check the RAM that stores the delays for the transmitted ultrasound
signals to the form focal beam.
Note:
You must reboot the system after performing the RAM test.
1) Access the
F.B.
tab of the
FE Diagnostics
dialog box, as described in the
Accessing the Front Board
Assembly Options
section, on page 7-25.
2) Trackball to the
RAM Test
button and press
Select
. The RAM test is performed and the results are
displayed in the
Comments
area, as shown in
Figure 7-24
(or
Figure 7-25
) below:
Figure 7-24 F.B. Tab - FE Diagnostics Dialog Box - RAM Test (RFI Configuration)