![GE Vivid 3 Service Manual Download Page 286](http://html.mh-extra.com/html/ge/vivid-3/vivid-3_service-manual_128704286.webp)
GE M
EDICAL
S
YSTEMS
D
IRECTION
2300164-100, R
EVISION
7
V
IVID
™ 3 P
RO
/V
IVID
™ 3 S
ERVICE
M
ANUAL
7-42
Section 7-3 - Performing Front End (FE) Diagnostics
Figure 7-23 Report Dialog Box - Pencil Probe Test (Close)
The
Report
dialog box includes the following:
•
Test Result Message:
Displays the test result.
•
Pencil Probe Relays:
Displays a graphic representation of the eight front board pencil probe
relays, as follows:
•
Each relay is represented by a button (highlighted in green) that displays the relay
number.
•
Each button is highlighted in yellow when selected.
•
Each button is highlighted in red if errors were detected for one or more relays.
•
Checkboxes:
Enable you to manipulate bad channels:
•
Show Only Bad Channels:
Select this checkbox to display only the bad channels in the
Display
area.
•
Mark bad Channel:
Select this checkbox to manually mark the currently selected vector
as bad.