Getting Started
1-34
Versana Essential
–
User Guide
5759238-100 English
Rev. 2
Connecting the System (continued)
Figure 1-18. Example Plug and Outlet Configurations
1. 100-120 VAC, 10A
Plug and Outlet Configuration
2. 220-240 VAC, 10A
Plug and Outlet Configuration
NOTE: Country-specific power cords are currently available for China,
India, Argentina, Europe, Japan, Switzerland, the United
Kingdom, the United States, Israel and Australia.
Voltage level check
Check the rating label on the rear side of the system. Check the
voltage range indicated on the label.
Connecting to the electrical outlet
WARNING
POWER OUTAGE MAY OCCUR. The ultrasound unit requires
a dedicated single branch circuit. To avoid circuit overload and
possible loss of critical care equipment, make sure you DO
NOT have other equipment operating on the same circuit.
CAUTION
Disconnect the plug from the wall outlet in case an emergency
should occur. Ensure easy access to the power outlet.