![GE Versana Essential Technical Publication Download Page 200](http://html.mh-extra.com/html/ge/versana-essential/versana-essential_technical-publication_128701200.webp)
After The Exam Is Over
3-60
Versana Essential
–
User Guide
5759238-100 English
Rev. 2
Setting up a Record Keeping System
Preparation
The following is needed:
• Quality
Assurance
binder.
•
Hard copy or electronic file of images.
• Quality
Assurance
Checklists.
•
Display the following information while testing quality
assurance:
• Acoustic
Output
• Gain
• Depth
• Probe
• Dynamic
Range
•
Set up new patient to be the name of the test.
•
Annotate the following:
•
Any control where its value is
NOT
displayed.
• Significant
phantom information.
Record Keeping
Complete the following:
1. Fill out the Ultrasound Quality Assurance Checklist for each
probe, as scheduled.
2. Make a hard copy or archive the image.
3. Compare images to baseline images and acceptable
values.
4. Evaluate trends over previous test periods.
5. File hard copy or electronic file of images and checklist in
Quality Assurance binder.