![GE Versana Essential Technical Publication Download Page 279](http://html.mh-extra.com/html/ge/versana-essential/versana-essential_technical-publication_128701279.webp)
Probe Discussion
Versana Essential
–
User Guide
5-23
5759238-100 English
Rev. 2
Specifications
Slice Thickness Specification
Table 5-8: System Probe Definitions
Probe Designation
Center Image
Frequency (MHz)
Frequency Range
(MHz)
Doppler Frequency
Range (MHz)
4C-RS
3.10 ± 10%
2.0~5.0
2.0~3.6
E8C-RS
6.5 ± 20%
4.2~10.0
4.2~6.3
3Sc-RS
2.75 ± 20%
1.7~4.0
1.7~3.3
C2-5-RS
3.5 ± 15%
2.0~5.0
2.0~3.6
L6-12-RS
7.75 ± 20%
4.0~13.0
4.0~6.0
L5-11-RS
7.5 ± 7%
3.7~12.0
3.7~5.6
Table 5-9: Slice Thickness Specification
Probe Designation
Probe Slice Thickness Specification [mm]
4C-RS
<= 12.0mm
E8C-RS
<= 13.0mm
3Sc-RS
<= 16.0mm
C2-5-RS
<= 13.0mm
L6-12-RS
<= 15.0mm
L5-11-RS
<= 8.0mm