Technical specifications
Operating data
200-700 nm, changeable wavelength filter units
Wavelength range
±2 nm
Filter center wavelength accuracy
< 10 nm
Filter bandwidth
< 5% deviation up to 2 AU at 280 nm with ferric
sulphate in 0.1 M sulphuric acid
Linearity
< 2x10
-4
AU at 280 nm
Short term noise (0.5–1 min)
1 2
< 2x10
-4
AU at 280 nm
Long term noise (1–10 min)
1, 2
< 2x10
-4
AU/h at 280 nm
Drift
2