![GE UVis-920 Operating Instructions Manual Download Page 50](http://html.mh-extra.com/html/ge/uvis-920/uvis-920_operating-instructions-manual_129559050.webp)
5.9
Check menu
Autozero
The instruments internal absorbance value for autozero can be checked to test the
consistency in buffers.
Display
Action
Step
Select menu
Check
, press
OK
.
1
Select menu
Autozero
. The autozero absorbance
value is shown.
2
Flow cell (checking flow cell type
and number)
This menu shows the type and serial number of the flow cell stored in the monitor.
Display
Action
Step
Select menu
Check
, press
OK
.
1
Select menu
Flow Cell
. If serial number is not set,
----- is shown. The serial value is shown number must
be 1000 or higher.
2
Check analog output
The function of the connected chart recorder can be tested.
Display
Action
Step
Select menu
Check
, press
OK
.
1
Select menu
Check Analog Out
, press
OK
.
2
50
Monitor UVis-920 Operating Instructions 29055049 AD
5 Operation
5.9 Check menu