3900/3900P User’s Manual
3-8
Condition
Possible cause(s)
Recommended action(s)
No communication
through the modem.
Incorrect cable used to connect
modem to monitor.
Modem is powered off or not
properly connected. Phone line
failure.
Incorrect or improperly entered
modem initialization string.
Your phone system may require a
guard tone.
Mismatched baud rate. For some
older modems, the baud rate for
the oximeter must be set to match
the baud rate for the modem.
Modem failure.
You must use the Datex-Ohmeda
modem/analog cable.
Refer to
Parts list
in chapter 6.
Check and correct all modem
connections. Power on the modem.
Check the phone line.
Check initialization string entry.
Verify the accuracy of the string. See
Modem setup
in chapter 5.
Enter a guard tone, if needed. See
Modem setup
in chapter 5.
Set the oximeter baud rate to match
the modem baud rate. See
SETTINGS
in chapter 2.
Contact the manufacturer of the
modem for assistance.
Data transmission from
the oximeter is very slow
or fails repeatedly.
The data transfer rate of the
modem connected to the oximeter
is too low.
Use a 56K modem.