B/Communications
B-5
RS-232 serial/analog interface cable—analog pinout
Use the Datex-Ohmeda
modem/analog
cable to connect an analog recording device
to the oximeter. Refer to
Parts list
in chapter 6.
The analog pinout is as follows:
RS 232
1
Pin 5 is used for signal ground
Pin 6 is used for analog output 1 (white jack, SpO
2
)
Pin 9 is used for analog output 2 (red jack, pulse rate)
Pins 1, 2, 3, 4, 7, and 8 are not used.
Connect the 9-pin connector on the modem/analog cable to the RS-232 connector
on the rear of the 3900/3900P, and connect the 3.5 mm white-cable jack (SpO
2
)
and the 3.5 mm red-cable jack (PR) to the recording device.
•
Jack tip carries the signal.
•
Jack base carries the signal ground.
Make sure the connections between the oximeter and the recording device are
secure.
To calibrate the recording device:
1.
On the oximeter’s front panel, press
and select SETTINGS.
2.
From the settings options, select ANALOG.
Use the +/– function key to cycle between the options (0.0V, 0.5V, and 1.0V).
3.
Select the voltage you want to calibrate on the recording device.
While at the settings screen, the selected voltage is then the only voltage
output through the analog channels.
4.
Verify on the recording device output that the voltage being sent is being
recorded correctly.
If it is not, follow the instructions that accompanied the recording device to
adjust it to the correct voltage.
5.
Repeat steps 3 and 4 to calibrate the other voltage.
6.
When calibration is complete, press
to return to the waveform screen.
Summary of Contents for TruTrak+ 3900
Page 1: ...GE Healthcare 3900 3900P Pulse Oximeter TruTrak User s Manual ...
Page 2: ......
Page 3: ...GE Healthcare 3900 3900P Pulse Oximeter TruTrak User s Manual 6050 0006 406 March 2005 ...
Page 26: ...3900 3900P User s Manual 1 16 ...
Page 80: ...3900 3900P User s Manual 5 18 ...
Page 96: ...A 4 3900 3900P User s Manual ...
Page 102: ...3900 3900P User s Manual B 6 ...
Page 106: ...Index ...