Section 5: Performance Verification and Safety Testing
2025653-048 Revision B
Responder
™
2000
Page 70
ECG to Serial 4.0 KVAC Test
1.
Setup the Hipot tester:
•
4.0 KVAC
•
High current limit 4.0 mA
•
Test 30 Sec.: Ramp 2.0 Sec.
•
120 VAC (±10 %) Input
2.
Plug in the ECG Connector Test Fixture into the ECG port.
3.
Plug the Serial Port Test Fixture into the Serial port.
4.
Connect the red hipot tester lead to the ECG Connector Test Fixture lead.
5.
Connect the black hipot tested lead to the Serial Port Test Fixture lead.
WARNING: Shock Hazard.
Before starting the test, ensure all lead connections are electrically isolated.
6.
Press
Start
(green button) on the Hipot tester. The test will take approximately 30 seconds.
•
If the test passes (short beep and green Pass LED is lit), continue with the next test.
•
If the test fails (audible alarm and red Fail LED is lit) discontinue testing. The unit must be returned to the
manufacturer for troubleshooting and repair.
AC Line/Neutral to AC Ground 1.5 KVAC Test
1.
Setup the Hipot tester:
•
1.5 KVAC
•
High current limit 4.0 mA
•
Test 30 Sec.: Ramp 2.0 Sec.
•
120 VAC (±10 %) Input
2.
Plug in the AC Plug Test Fixture to the AC input connector.
3.
Connect the red hipot tester lead to the AC Plug Test Fixture Line and Neutral lead.
4.
Connect the black hipot tested lead to the AC Plug Test Fixture Ground lead.
WARNING: Shock Hazard.
Before starting the test, ensure all lead connections are electrically isolated.
5.
Press
Start
(green button) on the Hipot tester. The test will take approximately two minutes.
•
If the test passes (short beep and green Pass LED is lit), continue with the next test.
•
If the test fails (audible alarm and red Fail LED is lit) discontinue testing. The unit must be returned to the
manufacturer for troubleshooting and repair.
ECG to SpO
2
1.5 KVAC Test
Note:
This test is only performed on models with the SpO
2
option installed.
1.
Setup the Hipot tester:
•
1.5 KVAC
•
High current limit 4.0 mA
•
Test 30 Sec.: Ramp 2.0 Sec.
•
120 VAC (±10 %) Input
2.
Plug in the ECG Connector Test Fixture into the ECG port.
3.
Plug the SpO
2
Test Fixture into the Serial port.
4.
Connect the red hipot tester lead to the SpO
2
Test Fixture lead.
5.
Connect the black hipot tested lead to Serial Port Test Fixture lead.
WARNING: Shock Hazard.
Before starting the test, ensure all lead connections are electrically isolated.
6.
Press
Start
(green button) on the Hipot tester. The test will take approximately two minutes.
•
If the test passes (short beep and green Pass LED is lit), continue with the next test.
•
If the test fails (audible alarm and red Fail LED is lit) discontinue testing. The unit must be returned to the
manufacturer for troubleshooting and repair.