SECTION 3: Troubleshooting
2025653-048 Revision B
Responder
™
2000
Page 24
Battery Troubleshooting
Check battery statistics. Good batteries have a capacity of at least 4500 mAH and will be charging when the Responder
2000 is plugged into an AC outlet.
1.
Ensure battery is fully charged.
2.
Perform Battery Calibration on page11.
3.
Replace the
Battery
.
4.
Replace
Power Control
Board
.
5.
Replace
Main
Board
.
SpO
2
Not Working
Verify the Responder 2000 has a SpO
2
Main
Board
installed.
Check the
About
screen. “
SpO2 - - - -
” indicates a non-SpO
2
Main
Board
is installed in the unit.
1.
Connect the Responder 2000 to a SpO
2
simulator with the appropriate sensor.
If the Responder 2000 SpO
2
display does not agree with the simulator setting, substitute a known good SpO
2
sensor.
2.
Check SPO
2
cable.
3.
Check cable from
Front Panel
to
SPO
2
Board
.
4.
Replace
SPO
2
Board
.
5.
Replace
Main
Board
.
Shock into Paddle Dock Fails
If shocking into the paddle dock does not work or shows high impedance.
Perform a button test for each Paddle.
•
If the test fails, replace the Paddle.
•
If the test does not fail, connect the Paddles to a simulator and attempt to shock.
If the shock is successful:
1.
Check the cables from the
Therapy
Board
to the paddle dock (one on each side).
2.
Replace the
Therapy
Board
.
If the shock is unsuccessful:
1.
Ensure the paddles are securely paced in the paddle dock.
2.
Replace the Paddles.
3.
Check the cables between the
Rear Panel
and the
Therapy Board
.
4.
Check the cable between the
Therapy Board
and the
Main Board
.
5.
Replace the
Therapy Board
.
6.
Replace the
Main Board
.