SECTION 3: Troubleshooting
2025653-048 Revision B
Responder
™
2000
Page 25
Shock into Patient Fails
1.
Perform a button test for each Paddle.
If the test fails, replace the Paddle.
If the test does not fail, connect the Paddles/Pads/Spoons to a simulator and attempt to shock.
2.
If the shock is successful:
The unit is working correctly. Check patient preparation.
3.
If the shock is unsuccessful:
1.
Replace the Paddles/Pads/Spoons.
2.
Check the cables between the
Rear
Panel
and the
Therapy
Board
.
3.
Check the cable between the
Therapy
Board
and the
Main Board
.
4.
Replace the
Therapy
Board
.
5.
Replace the
Main
Board
.
No ECG from Paddles/Pads/Spoons
1.
Replace the paddles/pads/spoons.
2.
Check the cables between the
Rear
Panel
and the
Therapy
Board
.
3.
Check the cable between the
Therapy
Board
and the
Main Board
.
4.
Replace the
Therapy
Board
.
5.
Replace the
Main
Board
.
Service Required LED is ON
If the Service Required LED is on and the fault has been corrected, perform this procedure to clear the fault.
1.
Go to
Maintenance
>
Stored Data Management
>
Clear Service Required Indicator
.
2.
Press
Accept
.
3.
Cycle power.
If the fault does not clear, go to
History
>
Event Log
to find the fault log entry.
Note:
All service required Error codes are displayed in the History with a red exclamation mark.