1. Phased Array Operation—Keypad, Menus, and Displays
PHASOR XS Operating Manual
Page 11
•
ANG CORRECTED
—Select straight down or angle corrected sector
scan (Refer to
Section 2.3.2
)
•
ASCAN MODE
—BUD (beam ultrasonic depth) setting causes the
A-Scan (if displayed) to be scaled to the depth of any reflector in the
selected beam (Refer to
Section 3.3
). PEAK setting is only available
with the optional TOP View (
Chapter 8
)
•
DATA SOURCE
—Selects amplitude or TOF data to appear in the
optional TOP View (
Chapter 8
)
IMAGE Submenu
(Refer to
Section 2.2
)
•
AMP COLOR PALET
—Adjusts the range of colors used to depict
the measured amplitude
•
TOF COLOR PALET
—Adjusts the range of colors used to depict
the time-of-flight (TOP View only,
Chapter 8
)
•
NEW DATA
—Indicate where “newest” data appears on screen as
it’s added during scrolling (TOP View only,
Chapter 8
)
•
VIDEO REVERSE
—Reverses sector, linear, or A scan’s orientation
BACKGRND Submenu
•
COLOR
—Changes the background and border colors of the display
(Refer to
Section 2.2
)
•
COLOR LEG
—Displays leg lines on Sector or Linear Scans to indi-
cate in which leg a reflector is located. (Refer to
Section 2.6.1
)
•
IMAGE BACKGRND
—Sets the display’s background color
•
A-SCAN COLOR
—Adjusts the color of the A-Scan (
Section 2.2
)
RESULTS1 Submenu
(Refer to
Section 3.2
)
•
READING 1 THROUGH READING 4
—Selects the measurement
displayed in each of the four small Reading Boxes.
RESULTS2 Submenu
•
LARGE READING
—Select the contents for display in the Large
Reading Box. (Refer to
Section 3.2
)
•
TTL #1
—Identifies in which gate an event triggers TTL 1 and illumi-
nates the Warning Light. (Refer to
Section 2.8.5
)
•
GATE SHAPE
—Limits A- or B-Gate functionality to detect only
those triggers that appear on the display or allows the gates to
detect all triggers that fall within the specified test-material range
(Refer to
Section 2.8.4
)
UT MENU
BASE Submenu
•
DISPLAY START
—Sets the display’s starting position to match the
Initial Pulse or IF gate trigger (for linear scans) or the point at which
sound enters the test material’s surface (all scans) (Refer to
Section 2.6.3
)
•
DISPLAY DELAY
—Shifts the Scan viewing window to ignore or
display initial material thickness (Refer to
Section 2.6.2
)
•
MATERIAL
—Selects the material being tested. Choose designa-
tions with “T” (transverse) for angle beam probes. Sets velocity to
value that approximately matches material specified. (Refer to
Section 2.4
)
•
LEG
—Sets the number of ultrasonic legs displayed, modifying the
displayed range. (Refer to
Section 2.6.1
)
PULSER Submenu
(Refer to
Section 2.7
)
•
VOLTAGE
—Sets pulser voltage level
•
WIDTH
—Sets width of square pulse
•
PRF VALUE
—Displays and/or allows adjustment of the Pulse Rep-
etition Frequency.
RECEIVER Submenu
(Refer to
Section 2.7
)
•
FREQUENCY
—Selects the frequency filter of the instrument.
•
ASCAN RECTIFY
—Selects the rectification-mode which effects how
the A-Scan appears on the display.
GATEMODE Submenu
•
GATE SELECT
—Select A, B, or IF gates (Refer to
Section 2.8
)
•
TOF MODE
—Indicates whether the triggering echo’s flank or
peak is used when making sound-path measurements. (Refer to
Section 2.8.2
)
•
LOGIC
—Determines whether the gate alarm is triggered when a
signal crosses the gate or does not cross the gate. (Refer to
Section 2.8.3
)
•
GATE DISPLAY
—Shows or hides selected gate. Gate continues to
function, even when hidden.
GATE POS Submenu
(Refer to
Section 2.8.1
)
•
GATE SELECT
—Select from A, B, or IF gates
•
GATE START
—Sets the beginning position of the selected gate.
•
GATE WIDTH
—Sets the width of the selected gate.
•
GATE THRESHOLD
—Sets the height of the selected gate as a
percentage of full-screen A-Scan height.
NRM/TCG Submenu
(Refer to
Section 3.6
)
•
TCG
—Turns on or off the effect of TCG
•
TCG DISPLAY
—Turns on or off the displayed TCG amplifier curve
•
CYC GAIN CNTL—
Used during TCG recording. Applies defined
gain slope across the entire range of cycles.
TCG EDIT Submenu
(Refer to
Section 3.6
)
•
BEAM CURSOR
—Select the beam to be edited
•
POINT
—Alter an existing point or create a new one
•
POINT POS
—Input or modify the time-based position of a TCG
point
•
POINT GAIN
—Input or modify the applied gain of a TCG point
FILES Menu
FILENAME Submenu
(Refer to
Sections 6.1
and
6.6
)
•
FILENAME
—Select stored files or input new data set or report
name.
•
SOURCE/DEST
—Indicates the device to or from which data is sent.
•
ACTION
—Recalls or deletes the selected file and saves edits to
data sets and reports.
•
ENTER
— Causes specified ACTION to occur.
HEADER Submenu
(Refer to
Section 6.5
)
•
HEADER NUMBER
—Selects the Header Line to edit.
•
EDIT
—Set to YES to allow editing of the selected Header Line.
•
HDR IN REPORT
—Determines if the header will be included in the
generated report.
MEMO Submenu
(Refer to
Section 6.4
)
•
EDIT
—Allows editing / creation of the Memo.
•
MEMO IN REPORT
—Determines if the memo will be included in the
generated report.
REPORT Submenu
(Refer to
Section 6.6
)
•
IMAGE IN REPORT
—Determines if the displayed Scan(s) will be
included in the generated report.
•
PARAM IN REPORT
—Determines if instrument settings will be
listed in the generated report.
CONFIG Menu
REGIONAL Submenu
(Refer to
Section 2.1
)
•
LANGUAGE
—Sets the language displayed on the instrument’s
screen
•
UNITS
—Sets displayed units to inch or millimeters
•
DECIMAL
—Select a period or comma for use as a decimal point.
•
DATE FORMAT
—How the date and time are displayed
STARTUP Submenu
•
STARTUP MODE
—Indicates if instrument is to start in the last
active mode or at the Welcome Screen (Refer to
Section 4.3
)
•
BRIGHTNESS
—Adjusts the display’s brightness (refer to
Section 2.2
)
•
DATE
—Sets the displayed date (Refer to
Section 2.1
)
•
TIME
—Sets the displayed time (Refer to
Section 2.1
)
Summary of Contents for Phasor CV
Page 6: ...Contents Page 4 PHASOR XS Operating Manual...
Page 16: ...1 Phased Array Operation Keypad Menus and Displays Page 14 PHASOR XS Operating Manual...
Page 24: ...2 Setting Up for Phased Array Measurement Page 22 PHASOR XS Operating Manual...
Page 86: ...5 Making Conventional Measurements Page 84 PHASOR XS Operating Manual...
Page 92: ...6 Storing Data Sets and Generating Reports Page 90 PHASOR XS Operating Manual...
Page 96: ...7 Technical Speci cations Page 94 PHASOR XS Operating Manual...