Statements
WARNING:
To avoid the risk of electrical shock, this equipment must only be connected to a supply
mains with protective earth.
The warmer does not adjust for patient temperature in warmup mode. When placing patient in the
unit, immediately change modes to manual mode or baby mode.
No additional tasks are required to power down the device after clinical use, other than turning off
the unit at the standby power switch on the front of the unit.
Isolation of the unit from the supply mains can only be achieved by turning off the device using the
mains power switch on the rear of the unit. The standby switch on the probe panel will not isolate the
device from the supply mains.
WARNING:
Electrical shock hazard: Before servicing, always unplug the unit from wall power.
User Responsibility
This Product will perform in conformity with the description thereof contained in this manual and
accompanying labels and/or inserts, when assembled, operated, maintained and repaired in
accordance with the instructions provided. This Product must be checked periodically. A defective
Product should not be used. Parts that are broken, missing, plainly worn, distorted or contaminated
should be replaced immediately. Should such repair or replacement become necessary, GE
Healthcare recommends that a telephone or written request for service advice be made to the
nearest GE Healthcare Regional Service Center. This Product or any of its parts should not be repaired
other than in accordance with written instructions provided by GE Healthcare and by GE Healthcare
trained personnel. The Product must not be altered without GE Healthcare’s prior written approval.
The user of this Product shall have the sole responsibility for any malfunction that results from
improper use, faulty maintenance, improper repair, damage or alteration by anyone other than GE
Healthcare.