B-1
Panda iRes Warmer
M1110737 006
© 2007 General Electric Company, all rights reserved
Appendix B
Specifications
__________________________________________
Power Requirements and Accessory Outlets
Power Requirements
Accessory Outlets
5.25 A @ 100v ~, 50/60 Hz
2 A @ 100v ~, 50/60 Hz
4.57 A @ 115v ~, 50/60 Hz
2 A @ 115v ~, 50/60 Hz
2.39 A @ 220v ~, 50/60 Hz
1A @ 220v ~, 50/60 Hz
2.28 A @ 230v ~, 50/60 Hz
1 A @ 230v ~, 50/60 Hz
2.19 A @ 240v ~, 50/60 Hz
1A @ 240v ~, 50/60 Hz
Standards
Designed to meet requirements of:
IEC 60601-2-21 with amendment
21 CFR CH-1 Section 1020.30 (n)
IEC 60601-1 with amendment
UL 60601-1
IEC 60601-1-2 with amendment
BSEN - 45501 with amendment
CSA/CAN C22.2 # 601.1
Operating Environment
Temperature
18 to 30°C
Humidity
5 to 75% Non-condensing relative humidity
Pressure
70 - 106 kPa
Air Velocity
up to 0.3 m/sec.
Storage conditions
Temperature
-25 to 60°C
Humidity
0 to 85% Non-condensing relative humidity
Pressure
50 to 106 kPa
User Control Settings
Patient control temperature 34-37.5°C in 0.1° increments
Radiant heat power 0-100% in 5% increments