4 Command and Query Summary
K0472 Revision A
4-58
:SENS:PRES:RES
Applicability:
PACE1000, PACE5000, PACE6000
Command Syntax
:SENSe[x][:PRESsure]:RESolution <string>
where x = 1 or 2 is the module number (default - 1)
Parameter:
<integer> resolution information
Short form:
:SENS:RES
Function:
Used to select a resolution to be used.
Query Syntax
:SENSe[x][:PRESsure]:RESolution?
where x = 1 or 2 is the module number (default - 1)
Short form:
:SENS:RES?
Function:
Asks for current resolution.
Response:
An integer from 4 to 7 representing selected resolution.
Description
This command selects the resolution to be used for showing the pressure reading in the front
panel display.
Example 1
TX> :SENS:PRES:RES?
RX> :SENS:PRES:RES 6
The resolution can be changed by the command, such as:
TX> :SENS:PRES:RES 4
Example 2
Query again confirms that the resolution has been changed to ‘4’ :
TX> :SENS:PRES:RES?
RX> :SENS:PRES:RES 4
Example 3
Any resolution number outside the 4 to 7 boundary generates error messages,
such as:
TX> :SENS2:PRES:RES 8
TX> :SYST:ERR?
RX> :SYST:ERR -222,"Data out of range; Parameter 1"
TX> :SENS2:PRES:RES 3
TX> :SYST:ERR?
RX> :SYST:ERR -222,"Data out of range; Parameter 1"