Table 8. 2450 Monitor Specifications
Conductivity Range
0-100 or 0-1000 (with proper sensor)
Conductivity Accuracy
±
1 or
±
10
µ
S
Conductivity Resolution
1 or 10
µ
S
pH Range
0-14 pH
pH Accuracy
±
0.05 pH
pH Resolution
0.01 pH
Deadband Adjustable
Setpoint
Direct or Reverse (configurable in the
field)
Keypad Numeric
Display
Illuminated 128x64 pixel LCD
Ambient Temperature
32-158
°
F (0-70
°
C)
Enclosure ABS
Plastic
A.4 Setup, Calibration and Customization
A.1.2
Check the Operation
Make sure the monitor has power and is operating. You will notice the monitor will display “SERIES
2000 REVERSE OSMOSIS MONITOR PRESS ANY KEY”. This indicates that power has been applied to the
monitor and no one has touched the keypad. This will also happen anytime there is a power outage
and power has been returned to the monitor.
1. Press any key on the keypad and you will see the Display Process Screen
2. Press CLR on the keypad and you will see the Main Menu Screen
3. Use the
↑
and
↓
arrow keys to move through the menu
A.1.3 Reinitialization
It is suggested that you reinitialize the monitor before programming in your own numbers. This will
wipe out any random settings which may be in the monitor. To reinitialize, follow these instructions.
1. After you have practiced moving up and down in the Main Menu, press 6 or highlight SYSTEM
SETUP and press ENT.
2. Press 2 or highlight INITIALIZATION and press ENT.
3. Press 2 or highlight WHOLE MONITOR and press ENT. A warning will appear on the screen
advising you that “this option requires recalibration and reprogramming!” Press 1 to proceed,
2 to cancel.
A.1.4
Set up Conductivity Preamp Settings and Temperature Compensation
Prior to testing your monitor, set up the internal conductivity preamp for the sensor you are using. If
you purchased a sensor with your monitor, the preamp is set up by the factory and you can proceed
directly to testing. If you change the sensor used and do not verify the preamp setup for your
attached sensor, the conductivity readings may be erroneous. Set up the preamp by using the
following steps.
Summary of Contents for Osmonics E4H Series
Page 36: ......
Page 39: ...E4H Series Operation Maintenance Manual 39 1 2450 Front Panel with Display ...
Page 45: ...E4H Series Operation Maintenance Manual 45 A 6 Technician Level Menus ...
Page 46: ...E4H Series Operation Maintenance Manual 46 A 1 8 Operation ...
Page 47: ...E4H Series Operation Maintenance Manual 47 ...
Page 48: ...E4H Series Operation Maintenance Manual 48 ...
Page 49: ...E4H Series Operation Maintenance Manual 49 A 1 9 Relays ...
Page 50: ...E4H Series Operation Maintenance Manual 50 A 1 10 Alarms ...
Page 51: ...E4H Series Operation Maintenance Manual 51 A 1 11 Flow Meters ...
Page 52: ...E4H Series Operation Maintenance Manual 52 A 1 12 4 20 mA Output ...
Page 53: ...E4H Series Operation Maintenance Manual 53 A 1 13 System Setup ...
Page 54: ...E4H Series Operation Maintenance Manual 54 ...
Page 55: ...E4H Series Operation Maintenance Manual 55 ...
Page 56: ...E4H Series Operation Maintenance Manual 56 ...
Page 57: ...E4H Series Operation Maintenance Manual 57 A 1 14 Clock ...
Page 58: ...E4H Series Operation Maintenance Manual 58 A 7 Security Level Menus ...
Page 59: ...E4H Series Operation Maintenance Manual 59 A 8 Operator View Level Menus ...