E4H Series
Operation & Maintenance Manual
25
TROUBLESHOOTING GUIDE
Symptom (Alarm)
Possible Cause
Remedies
Low operating
pressure
See the possible causes for low
operating pressure.
Dirty or fouled
membranes
Clean the membranes.
Operating on cold
water less than 55 -
60°F (13 - 16°C)
Install a hot/cold feed water
tempering valve if more permeate
flow is needed. Operate with a feed
water temperature of 72 - 77°F (22 -
25°C).
Membranes installed
backwards or
damaged concentrate
seal
Install membranes in the direction of
fluid flow. Clean the machine
immediately.
Membranes with damaged
concentrate seals should be
cleaned.
Flow meter inaccurate Check the flow rate manually with a
stopwatch and calibrated container
(such as CIP tank).
Low permeate flow
rate
Valve position
Check CIP and outlet valve positions
for correct orientation.
Concentrate valve
plugged
Disassemble and the clean the
plumbing to the valve.
Concentrate outlet line
restricted
Examine the concentrate line for
obstructions or a closed valve.
Low concentrate flow
rate, normal or higher
than normal pressure
Flow meter inaccurate Check the flow rate manually with a
stopwatch and calibrated container.
Note: CIP tank can be used for this
purpose.
Recycle or concentrate
lines plugged
Disassemble the plumbing to the
recycle orifice and remove foreign
particles.
Inaccurate pressure
gauge
Replace or calibrate the gauge as
required.
Service and CIP valves
closed at the same
time
Verify water path to both is open.
High operating
pressure
Restricted or reduced
permeate flow rate
See the possible causes for low
permeate flow rate.
Summary of Contents for Osmonics E4H Series
Page 36: ......
Page 39: ...E4H Series Operation Maintenance Manual 39 1 2450 Front Panel with Display ...
Page 45: ...E4H Series Operation Maintenance Manual 45 A 6 Technician Level Menus ...
Page 46: ...E4H Series Operation Maintenance Manual 46 A 1 8 Operation ...
Page 47: ...E4H Series Operation Maintenance Manual 47 ...
Page 48: ...E4H Series Operation Maintenance Manual 48 ...
Page 49: ...E4H Series Operation Maintenance Manual 49 A 1 9 Relays ...
Page 50: ...E4H Series Operation Maintenance Manual 50 A 1 10 Alarms ...
Page 51: ...E4H Series Operation Maintenance Manual 51 A 1 11 Flow Meters ...
Page 52: ...E4H Series Operation Maintenance Manual 52 A 1 12 4 20 mA Output ...
Page 53: ...E4H Series Operation Maintenance Manual 53 A 1 13 System Setup ...
Page 54: ...E4H Series Operation Maintenance Manual 54 ...
Page 55: ...E4H Series Operation Maintenance Manual 55 ...
Page 56: ...E4H Series Operation Maintenance Manual 56 ...
Page 57: ...E4H Series Operation Maintenance Manual 57 A 1 14 Clock ...
Page 58: ...E4H Series Operation Maintenance Manual 58 A 7 Security Level Menus ...
Page 59: ...E4H Series Operation Maintenance Manual 59 A 8 Operator View Level Menus ...