Function:
1. Cabinet monitor will trip the E-Off contactor in the PDU when
a) Leak Sensor 2 detects water leakage or are open
b) T
emperature of system cabinet is over 45 degree centigrade
c) 4KW LCS or 8KW LCS/MCS stop working
2. Cabinet monitor only sends the error message to host through CAN link when
a) Leak Sensor 3 detects water leakage or are open
b) T
emperature of system cabinet is over 40 and less 45 degree centigrade
c) Water level is low in any one of LCS or MCS
3. Cabinet monitor will trip the Night Mode contactor in the PDU when
The system haven
’
t scanned for one hour.
The Nigh Mode control signal is from IRF I/O board J23 to SCIF J101, then to SCIF J105, then to J10
in cabinet monitor.
4. Cabinet monitor will
energize
the Night Mode contactor in the PDU when
The system is running TPS reset or begins to scan.
5. Provide Is24V DC power to whole CAN link.
Pump flow sensor and Tank level sensor are all Normally Closed switch, which means:
1. When LCS or MCS stop working, the pump flow sensor are close. And when they works normally,
the sensors are open.
2. When Tank level is low, the tank level sensor are close. And when the level is normal, the sensors
are open.
You can use DVM to check the resistance between the two sensor cables.
Summary of Contents for OPTIMA MR360
Page 3: ...1...
Page 4: ...2...
Page 5: ...3...
Page 7: ...5...
Page 8: ...6...
Page 11: ...9 The portable GasWatch2 Oxygen sensor as pictured above part number is 5111049...
Page 14: ...12...
Page 15: ...13...
Page 16: ...14...
Page 17: ...15...
Page 18: ...16...
Page 20: ...18...
Page 21: ...19...
Page 33: ...31...
Page 38: ...36...
Page 39: ...37...
Page 41: ...39...
Page 43: ...41...
Page 44: ...42...
Page 45: ...43...
Page 47: ...45...
Page 48: ...46...
Page 49: ...47...
Page 50: ...48...
Page 51: ...49...
Page 52: ...50...
Page 53: ...51...
Page 54: ...52...
Page 55: ...53...
Page 56: ...54...
Page 62: ...60...
Page 65: ...63...
Page 66: ...64...
Page 67: ...65...
Page 68: ...66...
Page 69: ...67...
Page 71: ...69...
Page 78: ...Image can be previewed in this new version of Image Management 76...
Page 79: ...77...
Page 80: ...78...
Page 81: ...79...
Page 82: ...80...
Page 85: ...83...
Page 88: ...86...
Page 89: ...87...
Page 90: ...88...
Page 91: ...89...
Page 96: ...94...
Page 97: ...95...
Page 102: ...100 This slide shows the detailed configuration of THTF host...
Page 104: ...102...
Page 105: ...103...
Page 106: ...104...
Page 112: ...110...
Page 116: ...114...
Page 117: ...115...
Page 124: ...122...
Page 129: ...127...
Page 131: ...129...
Page 134: ...132 The part is the 200V DC power supply for three XFAs and 200V to 700V DC DC power supply...
Page 135: ...133...
Page 136: ...134 Three sets of DC700V and 700V are isolated...
Page 142: ...140...
Page 143: ...141...
Page 147: ...145...
Page 150: ...148...
Page 154: ...152...
Page 170: ...168...
Page 171: ...169...
Page 172: ...170...
Page 174: ...172...
Page 176: ...174...
Page 180: ...178...
Page 185: ...183...
Page 186: ...184...
Page 187: ...185...
Page 188: ...186 The report file is stored in export home signa bin...
Page 189: ...187...
Page 190: ...188...
Page 192: ...190 The RF and DC cables for Express PA Coil Cradle mode are shipped with the system...
Page 193: ...191...
Page 220: ...Page 2 System Cabinet 1 2 3 4 5 6 2 3 4 5 6 1...
Page 221: ...Page 3 System Cabinet 1 2 3 4 5 6 1 3 4 2 5 6...
Page 222: ...Page 4 System Cabinet 1 2 3 4 5 6 4 5 6 2 3 1...
Page 223: ...Page 5 System Cabinet 1 2 3 4 5 6 7 8 9 10 11 2 3 4 5 6 7 8 9 10 11 1...
Page 224: ...Page 6 Magnet Side Kit 1 2 3 1 2 3...
Page 225: ...Page 7 Rear Pedestal 1 2 3 4 5 6 7 4 3 1 2 5 6 7...
Page 226: ...Page 8 LPCA 1 2 1 2...