157
There are five limit switches in the Fixed table.
1. One Home Sensor Limit Switch. Only when cradle moves to the home position, the Home
sensor Limit Switch will be activated to allow the fixed table to be moved up and down. It
can only be accessed for service by removing the top right cover.
2. Two Up Limit Switches and Two Down Limit Switches. (
Shows the logic position diagram of
the four limit switch
) This diagram shows the logical position relationship of the four limit
switches. Only Up1 Limit Switch locates on the fixed table frame. The other three limit
switches integrate in the actuator. When the table is going down, Down1 LS will be
touched first, then fixed table will stop if Down1 LS is OK. But if the Down1 LS doesn
’
t
work, the table will keep going down, Down2 LS will be touched and the table will stop. It
is same for Up1 LS and Up2 LS. In normal situation, only Up1 and Down1 will be touched.
Up2 and Down2 LS are redundant to assure the table can stop at UP and DOWN limit
position in case Up1 or Down1 LS are broken. The highest position of the fixed table can
be determined by adjusting the position of the UP LS Actuator. Only when the fixed table
is on the top height, the cradle can be moved in and out.
Summary of Contents for OPTIMA MR360
Page 3: ...1...
Page 4: ...2...
Page 5: ...3...
Page 7: ...5...
Page 8: ...6...
Page 11: ...9 The portable GasWatch2 Oxygen sensor as pictured above part number is 5111049...
Page 14: ...12...
Page 15: ...13...
Page 16: ...14...
Page 17: ...15...
Page 18: ...16...
Page 20: ...18...
Page 21: ...19...
Page 33: ...31...
Page 38: ...36...
Page 39: ...37...
Page 41: ...39...
Page 43: ...41...
Page 44: ...42...
Page 45: ...43...
Page 47: ...45...
Page 48: ...46...
Page 49: ...47...
Page 50: ...48...
Page 51: ...49...
Page 52: ...50...
Page 53: ...51...
Page 54: ...52...
Page 55: ...53...
Page 56: ...54...
Page 62: ...60...
Page 65: ...63...
Page 66: ...64...
Page 67: ...65...
Page 68: ...66...
Page 69: ...67...
Page 71: ...69...
Page 78: ...Image can be previewed in this new version of Image Management 76...
Page 79: ...77...
Page 80: ...78...
Page 81: ...79...
Page 82: ...80...
Page 85: ...83...
Page 88: ...86...
Page 89: ...87...
Page 90: ...88...
Page 91: ...89...
Page 96: ...94...
Page 97: ...95...
Page 102: ...100 This slide shows the detailed configuration of THTF host...
Page 104: ...102...
Page 105: ...103...
Page 106: ...104...
Page 112: ...110...
Page 116: ...114...
Page 117: ...115...
Page 124: ...122...
Page 129: ...127...
Page 131: ...129...
Page 134: ...132 The part is the 200V DC power supply for three XFAs and 200V to 700V DC DC power supply...
Page 135: ...133...
Page 136: ...134 Three sets of DC700V and 700V are isolated...
Page 142: ...140...
Page 143: ...141...
Page 147: ...145...
Page 150: ...148...
Page 154: ...152...
Page 170: ...168...
Page 171: ...169...
Page 172: ...170...
Page 174: ...172...
Page 176: ...174...
Page 180: ...178...
Page 185: ...183...
Page 186: ...184...
Page 187: ...185...
Page 188: ...186 The report file is stored in export home signa bin...
Page 189: ...187...
Page 190: ...188...
Page 192: ...190 The RF and DC cables for Express PA Coil Cradle mode are shipped with the system...
Page 193: ...191...
Page 220: ...Page 2 System Cabinet 1 2 3 4 5 6 2 3 4 5 6 1...
Page 221: ...Page 3 System Cabinet 1 2 3 4 5 6 1 3 4 2 5 6...
Page 222: ...Page 4 System Cabinet 1 2 3 4 5 6 4 5 6 2 3 1...
Page 223: ...Page 5 System Cabinet 1 2 3 4 5 6 7 8 9 10 11 2 3 4 5 6 7 8 9 10 11 1...
Page 224: ...Page 6 Magnet Side Kit 1 2 3 1 2 3...
Page 225: ...Page 7 Rear Pedestal 1 2 3 4 5 6 7 4 3 1 2 5 6 7...
Page 226: ...Page 8 LPCA 1 2 1 2...