2 Altitude and Atmospheric Pressure
Refer to
. To obtain altitude and atmospheric pressure requirements for components
not specified in
, refer to the appropriate component Pre–Installation Manual listed in
Chapter 2, Basic System Compatibility
Table 4-2: Altitude and Atmospheric Pressure
INSTALLATION ROOM OF PRODUCT OR
COMPONENT
ALTITUDE (meters)
ATMOSPHERIC PRESSURE (kPa)
IN-USE
IN-USE
MIN
MAX
MIN
MAX
Examination room
0
2000
79.4
106
Technical room (See Notice below
0
2000
79.4
106
Control room
0
2000
N/A
N/A
NOTICE
The chiller is able to dissipate maximum continuously power at 3000 meters in an
ambient temperature up to 20°C, for this altitude the technical room temperature
shall not exceed 20°C.
Optima IGS 320, Optima IGS 330 Pre-Installation Manual
Direction 5537562-1-1EN, Revision 3
Chapter 4 Environmental Requirements
157