3.3 Mounting Requirements
3.3.1 Floor Loading and Recommended Mounting Methods
. To obtain floor loading and recommended mounting methods for components
, refer to the appropriate component Pre-Installation Manual listed in
.
Table 2-7:
PRODUCT OR
COMPONENT
NET WEIGHT
KG (LBS)
DIMENSIONS MM (INCHES)
LOAD BEAR‐
ING AREA
MM (IN‐
CHES)
WEIGHT/ OC‐
CUPIED
AREA
MOUNTING METHOD
WIDTH
DEPTH
HEIGHT
LC Positioner
750 (1655)
See Illustrations
LC POSITIONER DI‐
MENSIONS
:
•
Side view
•
Top view
•
Front view
in
Circle diame‐
ter 600
(23.62)
Recommended:
•
Through-Bolts
(12)
Alternates:
•
On Grade 5/8 in,
Anchors (12)
•
Above Grade 3/4
in. Anchors (12)
See
Table
590 (1300)
See note 1
See illustrations in
•
Omega IV
•
Omega V
571.5x429
(22.5x16.9)
2410 kg/m
2
(492.3 lb/ft
2
)
Same as LC Position‐
er See IMPORTANT
NOTICE below.
System Control
Cabinet
See Illustration
System Control Cabi‐
net Dimensions
in
1206x717(47.
5x28.2)
740 kg/m
2
(152 lb/ft
2
)
COOLIX 4100
120 (264.5)
555 (21.8)
610 (24) 1200 (47.2)
Autotransformer
(Coolix 4100)
30 (66)
370 (14.5)
304 (12)
340 (13.4)
Detector Chiller
Thermo-Con
14.6 ( 32.2)
See Illustration
Detector Chiller Ther‐
mo–Con Dimensions
344x283
(13.5x11.14)
Fluoro UPS UL
(optional)
530 (1169)
690 (27.0) 800 (31.5) 1820 (70.1)
975 kg/m
2
(200 lb/ft
2
)
Ground -mounted
Fluoro UPS CE
(optional)
480 (1059)
680 (26.7) 800 (31.5) 1450 (57.1)
883 kg/m
2
(181 lb/ft
2
)
Ground -mounted
Fluoro UPS IF box
(optional)
4 (9)
280 (11.02)
314
(12.36)
124.5 (4.9)
Wall-mounted
PDB US (4)
148 (326)
685 (27)
225 (9)
1850 (73)
Wall-mounted
PDB CE (4)
110 (242)
800
300
1200
Wall-mounted
EMI Filter enclo‐
sure CE (4)
<40
300
210
950
Wall-mounted
DL keypad
1.4 (3)
283 (11.55) 300 (11.8)
82 (3.25)
Optima IGS 320, Optima IGS 330 Pre-Installation Manual
Direction 5537562-1-1EN, Revision 3
Chapter 2 Equipment Requirements
113