CAUTION
The System delivered with the Open Monitor Suspension option cannot presume
on the mechanical constraints of non-GE monitor(s) suspension(s) introduced in
the system.
For further details on Open monitor suspension Pre-installation, please refer to Regulatory
Requirements and Pre-Installation Instructions contained in Open Monitor Suspensions -
Service Instruction for Installation, 5691988-1-1EN contained in the Service manual.
1.2.8 Description of the Ergo monitors suspensions
1.2.8.1 Suspension System Configuration
The monitor suspensions are mainly supplied with any vascular system; it enhances the X–Ray
functions by adding the ability to show the image monitor in Examination Room.
Ergo LCD suspension should be mounted to one pair of stationary rails, which include the cable
tracks for handling cables of suspension.
Illustration 2-7: Ergo LCD suspension
Optima IGS 320, Optima IGS 330 Pre-Installation Manual
Direction 5537562-1-1EN, Revision 3
Chapter 2 Equipment Requirements
51