OEC 9800 Installation Procedure
30
4.
Position the C-arm so the II is closest to the floor.
5.
Complete a calibration by performing the following.
A
Connect the Transmitter, and both receivers (Probe and Calibration Fixture) to the panel located on the left side of the
Workstation.
C. Place the tip of the Probe in the dimple located on the Transmitter.
PROBE
TRANSMITTER
TRANSMITTER
DIMPLE
Figure 13. Probe Touching Dimple on Transmitter
D. Press the CALIBRATE button on the FPD touch screen.
Note
:
When the calibration is accepted the CALIBRATE button on the Visualization menu changes from orange to green.