OEC 9800 Installation Procedure
27
Digital Hardcopy Camera test
1.
Verify operation of the digital hardcopy camera. Refer to the Operator's Guide Supplement if necessary.
Instant Film/Paper (Onboard) Printer
1.
Remove the shipping pins from the printers front ventilation grill. There are two pins connected with tape which secure the
thermal printhead.
2.
Verify that an image can be printed to the Instant Film/Paper Printer. Refer to the Operator's Guide Supplement if necessary.
Thermal Printer Test
1.
Verify an image can be printed to the thermal printer.
DICOM
1.
If DICOM option is available: Contact the system administrator and ask them if they are ready to have the DICOM option
installed.
If they are ready, have the system administrator complete the DICOM Network Questionnaire found as the last section of this
document (Appendix 2), then attach it to the
9800 Installation Report/Checklist
and mail them in together. Then Continue
with DICOM installation.
If the Hospital is not ready for DICOM installation: photocopy the DICOM Network Questionnaire, leave it with the system
administrator or equivalent, and schedule a time to complete DICOM installation. In the
Report/Checklist
, record the
scheduled time that DICOM is to be installed (indicate time by MMDDYY). Continue with system installation.
Note
:
If the System Administrator completed a “copy” of the Network Questionnaire, attach the copy to the
9800 Installation
Report/Checklist
and mail them in together.
2.
Make electrical connections and configure the Workstation software.