Initial Setup
2-15
April 2004
Entering Delta F Cell
Calibration Data (cont.)
5. Move the pointer to the Zero
µ
A line and press [YES].
6. Enter the microamp (
µ
A) value and press [YES].
7. Move the pointer to the Zero ppm (or %) line and press [YES].
8. Enter the zero range value, and press [YES]. The range units will
be either ppm, ppb or %, depending on the cell type/model
selected in the Probe Configuration Menu.
9. Repeat Steps 5 through 8 to enter the microamp (
µ
A) and range
value for the span line.
IMPORTANT:
Leave the O2 Current Multiplier at 1.00 unless you
are using a background gas other than nitrogen. See
the section Background Gas Correction Factors for
the Delta F Oxygen Cell in Chapter 2,
Troubleshooting and Maintenance, of the Service
Manual.
Repeat Steps 1 through 9 to enter oxygen calibration data for the
remaining channels.
To enter calibration data for moisture probes, refer to A. Entering
Moisture Probe Calibration Data, page 2-12. (For other types of
probes, see Chapter 2 of the Programming Manual.) To exit, press the
DONE menu key until Main Menu appears on the message line. Then
refer to Entering High and Low Reference Values on page 2-16.
Figure 2-7: System Calibration Menu for Oxygen Cell
Oxygen Probe Calibration Ch
System Calibration Menu
HELP CHANNEL PROBE DONE
S/N
µA ppm O
2
Zero: ___ ___
Span: ___ ___
O2 Current Multiplier: 1:00
1