April 2004
3-12
Specifications
Delta F Oxygen Cell
Type
Non-depleting electrolytic oxygen sensing cell.
Available Cells
PPB
v
O
2
Range:
L: 0 to 500 ppb
v
/5 ppm
v
/50 ppm
v
Ranges for each cell are software selectable in GE Panametrics
analyzers.
PPM
v
O
2
Ranges:
A: 0 to 1/10/100 ppm
v
B: 0 to 10/100/1000 ppm
v
C: 0 to 100/1000/10,000 ppm
v
D: 0 to 50/500/5000 ppm
v
Ranges for each cell are software selectable in GE Panametrics
analyzers.
Percent O
2
Ranges:
A: 0 to 5%
B: 0 to 10%
C: 0 to 25%
Accuracy
±1% full scale (ranges > 0 to 2.5 ppm
v
)
±5% full scale (ranges < 0 to 2.5 ppm
v
)
Sensitivity
< 5 ppb (0 to 500 ppb
v
range)
Response Time
Sensor responds instantaneously to O
2
change.
Equilibrium time is application specific.
Ambient Temperature
0 to 49°C (32 to 120°F)
Background Gas
Compatibility
Standard Cell:
Ultra-pure inert gases
STAB-EL™ Cell:
All gas compositions including those containing “acid” gases such as
CO
2
, H
2
S, Cl
2
, NO
x
, SO
2
, etc.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com