Troubleshooting and Maintenance
2-9
June 2003
Procedure for Trimming
the Span Value
1.
Use the arrow keys to move the pointer to the recorder you want to
test and press [YES].
2.
Enter 100 for % of scale and press [YES].
3.
Observe the multimeter reading. Wait at least 5 seconds for the
recorder output to settle. The meter should display one of the
following:
4.
Press the TRIM menu key.The Series 1 displays a recorder trim
value next to the 100% setting. The trim value will be labeled Zero
or Span. If the Zero value displays, press the ZRO/SPAN menu
key to display the Span trim value.
5.
Press the TRIM UP or TRIM DN menu key to set the trim span
value to the difference between the desired and actual multimeter
readings in Step 3. Confirm that the digital multimeter display
reads the correct 100% of full scale value (20 mA or 2V).
To exit, press the DONE menu key until Main Menu appears on the
message line.
Note:
The zero trim is an offset adjustment, while the span trim is a
slope adjustment. As a result, the zero and span trim affect
each other. Therefore, after you adjust one, you may have to
adjust the other.
For the Recorder Output
Range
Desired Multimeter Reading
0 to 20 mA
20 mA
4 to 20 mA
20 mA
0 to 2 V
2 V