June 2003
1-6
Installing Optional Features
Connecting the Alarms
You can order the Series 1 with high and low alarm relays for each
channel. Hermetically sealed alarm relays are also available. Each
alarm relay is a single-pole double-throw contact set that contains the
following contacts (see Figure 1-4):
•
normally open (NO)
•
normally closed (NC)
•
armature contacts (C)
Make connections for the high and low alarm relays on the desired
channel(s) on the terminal blocks labeled ALM A and ALM B on the
back panel of the electronics unit. Use Table 1-3 to make high and
low alarm connections. See Figure 1-5 for the terminal block
locations.
Note:
The alarm terminal block has an addition Return connection
that you can use to ground the alarms if desired.
Figure 1-4: Alarm Relay Contact Points
Table 1-3: Alarm Connections
Connect Low Alarm:
To ALM A Terminal Block:
NO contact
pin NO
C contact
pin C
NC contact
pin NC
Connect High Alarm:
To ALM B Terminal Block:
NO contact
pin NO
C contact
pin C
NC contact
pin NC
NO
C
NC