MG-101 User’s Manual
13
Chapter 2. Setup and Operation
2.2.5 Preparation for Shipping
To ship or to move the MG-101, refer to
below and empty the saturator bottle as follows:
1.
Remove the rear panel by removing the two small screws at the bottom.
2.
Fully open the vent valve by turning it counter-clockwise.
3.
Place the end of the fill tubing in a container below the level of the saturator bottle and open the petcock.
4.
When the saturator bottle is completely empty, close the petcock and the vent valve, and reinstall the rear panel.
Figure 7: Emptying the Saturator Bottle
Saturator Bottle
Fill Tubing
Petcock
Screws (2)