Marquette Hellige GmbH
MAC 1100/1200 V 1.1
Page
57
227 492 20 D - 0002
- During single fault conditions (S.F.C.), referring to the electrical diagram, the
measurements have to be done under the following conditions:
* Polarity switch
NORM and RVS
* GND switch
GND open
* S1
closed
Test has failed if the measured values are greater than:
N.C.
S.F.C
100
µ
A 500
µ
A
300
µ
A (UL requirements)
Electrical Diagram for Enclosure Leakage Current Test
7.1.3.3.2 Patient Leakage Current Test
This test performs a leakage current test
under single fault conditions (S.F.C.) de-
pending on domestic power outlet of 115
or 230 V AC as source into the floating
inputs.
In all cases, the leakage current is
measured from input jack of the unit
being tested to ground.
Connect the unit being tested to your Safety Tester.
Referring to the electrical diagram, measurements have to be done under the fol-
lowing conditions:
* Polarity switch NORM and RVS
* GND switch GND closed
*
S1
closed
Test has failed if the measured values are greater than 50
µ
A.
Summary of Contents for marquette MAC 1100
Page 1: ...MAC 1100 1200 Servicing Instructions Version 1 1 227 492 20 SA e Revision D...
Page 6: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 6 227 492 20 D 0002...
Page 40: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 40 227 492 20 D 0002...
Page 44: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 44 227 492 20 D 0002...
Page 48: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 48 227 492 20 D 0002...
Page 50: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 50 227 492 20 D 0002...
Page 62: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 62 227 492 20 D 0002...
Page 70: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 70 227 492 20 D 0002...
Page 72: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 72 227 492 20 D 0002...
Page 73: ......
Page 74: ......
Page 75: ......
Page 86: ......
Page 87: ......
Page 89: ......
Page 91: ......