3.10 Time and date
This section deals with the quick checking and setting of the time and date.
Pressing the
“7"
key calls up the following display:
Time
(hh:mm)
[18:25:29]
and
Date
(dd.mm.yyyy)
[03.03.1999]
X = End
To set time or date, select the corresponding time or date field by pressing the return
key. During entering the numbers, a single cursor is not displayed.
3.11 Electrode test
The electrode test is started by pressing the
“8" key.
The following menu is displayed:
Electrode test
Terminate with any key
Select pace electrode (key ‘P’)
Pace electrode LA
Cable = 10 electrodes (or no cable)
The status of each electrode is displayed. Electrode OK or disconnected is displayed
as follows:
Status = 0 -> Electrode OK
Status = 1 -> Electrode disconnected
With the “P” key, the pace electrode can be selected. The selected pace electrode is
displayed.
The status
“cable”
indicates if a cable is connected to the patient ECG input or not.
Summary of Contents for marquette MAC 1100
Page 1: ...MAC 1100 1200 Servicing Instructions Version 1 1 227 492 20 SA e Revision D...
Page 6: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 6 227 492 20 D 0002...
Page 40: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 40 227 492 20 D 0002...
Page 44: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 44 227 492 20 D 0002...
Page 48: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 48 227 492 20 D 0002...
Page 50: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 50 227 492 20 D 0002...
Page 62: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 62 227 492 20 D 0002...
Page 70: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 70 227 492 20 D 0002...
Page 72: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 72 227 492 20 D 0002...
Page 73: ......
Page 74: ......
Page 75: ......
Page 86: ......
Page 87: ......
Page 89: ......
Page 91: ......