Paper transport functions, no printout
-
Paper feed flap correctly engaged on both sides?
-
All connectors for printhead plugged in?
-
Operation with strongly depleted battery?
if yes:
-
plug in power plug
if no:
-
PCB Control defective
- or printhead defective
MAC 1200 only prints on the upper or lower section of the printout
- Paper feed flap is only engaged on one side.
MAC 1200 prints, but only baselines are printed out
-
Electrodes applied correctly?
-
Electrode cables plugged into the patient trunk cable terminal-box correctly?
-
Patient trunk cable defective ( e.g., RL defective)?
-
Contact problems at the patient input connector of the MAC1200?
if not:
PCB Control defective.
Summary of Contents for marquette MAC 1100
Page 1: ...MAC 1100 1200 Servicing Instructions Version 1 1 227 492 20 SA e Revision D...
Page 6: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 6 227 492 20 D 0002...
Page 40: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 40 227 492 20 D 0002...
Page 44: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 44 227 492 20 D 0002...
Page 48: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 48 227 492 20 D 0002...
Page 50: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 50 227 492 20 D 0002...
Page 62: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 62 227 492 20 D 0002...
Page 70: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 70 227 492 20 D 0002...
Page 72: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 72 227 492 20 D 0002...
Page 73: ......
Page 74: ......
Page 75: ......
Page 86: ......
Page 87: ......
Page 89: ......
Page 91: ......