3.12 Time constant
By pressing the
“9"
key, the time constant for the ECG patient input can be selected.
The following menu is displayed
1=
1.02s
( frequency
0,16Hz )
2=
2.04s
( frequency
0.08Hz )
3=
4.08s
( frequency
0,04Hz )
X = End
The currently selected time constant appears in reverse video. Use key 1 ... 3 for
selection.
The factory setting is 2.04s. This should only be changed when deemed really
necessary..
3.13 Setting the device model
The setting of the device model is activated by pressing the
“D” key
, the following
display image appears:
D e v I c e model
1=
INT
2=
USA
3=
Europe 2
4=
ASIA
(not available from version V1.1)
5=
RUSS
X=
End
Warning!
Changing the device model affects the selection of the output formats and
languages available, can switch the program for Interpretation from HEART to 12
SL, or vice versa. Also affected are the default configuration settings, e.g.,
override function yes/no, line filter 50/60 Hz.
The device model configured during manufacture should only be changed when
deemed really necessary.
Summary of Contents for marquette MAC 1100
Page 1: ...MAC 1100 1200 Servicing Instructions Version 1 1 227 492 20 SA e Revision D...
Page 6: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 6 227 492 20 D 0002...
Page 40: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 40 227 492 20 D 0002...
Page 44: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 44 227 492 20 D 0002...
Page 48: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 48 227 492 20 D 0002...
Page 50: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 50 227 492 20 D 0002...
Page 62: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 62 227 492 20 D 0002...
Page 70: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 70 227 492 20 D 0002...
Page 72: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 72 227 492 20 D 0002...
Page 73: ......
Page 74: ......
Page 75: ......
Page 86: ......
Page 87: ......
Page 89: ......
Page 91: ......