GE LightSpeed 16 Product Data Sheet Download Page 6

g

 

Product Data Sheet – rev 4.2 
July 4th 2003  
Page 6 
 
 

 

LightSpeed

16

 

 

With Xtream 

CT Scanner System 

GE Medical Systems-America 
Milwaukee, USA  - Fax: 1 414 544 3384 
GE Medical Systems-Asia: 
Tokyo, Japan - Fax: 81 425 85 5490 
Hong Kong - Fax: 852 2559 3588 
GE Medical Systems-Europe 

next available recon. Available during or 
after scanning. 

Total Scan 

Total Elapsed 

Max 

Time (sec.)  Time (sec.) 

mA 

Helical Scan Protocols 

Images annotated to indicate continuous 
scan acquisition with table incrementation: 

150 155 

100-200 

200 205 

80-150 

450 465 

50-70 

All protocols assume 120 kVp scans under 
typical clinical conditions.

  

HE (helical) + Pitch 

Single Helical Scans: 

Helical Scan Image Quality 

Table speed 

 

Prospective Multiple Reconstruction

 

(PMR): Up to 3 sets of reconstructions can 
be pre-programmed as part of the scan 
protocol prior to acquisition.  The operator 
can select different start/end location, slice 
thickness, interval, reconstruction 
algorithms and display fields of view for 
each reconstruction.  This frees the 
operator from sitting at the console and 
directly contributes to increased 
productivity. 

Scan  
Time 

Maximum 
mA 

10 sec. 

440 

20 390-440 
30 330-380 
40 290-350 
50 260-320 
60 230-310 
70 210-300 
80 180-280 
90 170-270 
100 160-260 
110 150-250 
120 140-230 

Volumetric Helical Scan Image Quality 

With LightSpeed

16

 being a sub-millimeter 

isotropic CT scanner, it is now possible to 
specify coronal and sagittal image quality: 

1. Visual 

Measurement: 

Reformatted resolution is demonstrated on 
the Nuclear Associates AAPM High 
Contrast Resolution Insert #76-413. 

Prospective Recon

: Operator may initiate 

full recons at any table location in 
increments of 1/10 the image thickness; 
image thickness remains constant.  

          0.5mm 

 

Retrospective Recon:

 Operator may 

initiate full recons at any table location in 
0.1 mm increments; image thickness 
remains constant. 

Retrospective Image Decomposition

The operator has the option to 
retrospectively decompose the original 
raw data set and reconstruct additional 
images at any of the defined nominal 
image thickness available for a given table 
speed and scan mode. 

 

Multiple Helical Scans  

 

(contiguous helical coverage with 5-
second IGD):

 

 

Scan

Time

IGD

No. 

Scans

Max

mA

10 sec.

5 sec.

3

340-420

4

300-400

5

270-390

6

240-360

20 sec.

5 sec.

2

300-360

3

240-330

4

180-310

30 sec.

5 sec.

2

230-310

3

180-280

4

140-240

40 sec.

5 sec.

2

190-290

3

140-230

4

110-190

50 sec.

5 sec.

2

170-270

3

120-200

4

90-160

60 sec.

5 sec.

2

140-240

3

100-170

 

 

     0.4mm 

An Effective 0.5 mm voxel size is clearly 
seen in coronal and sagittal views. 

2. Statistic 

Measurement: 

Volumetric MTF is computed from the X,Y 
and Z axis of a 0.2mm bead using sagittal 
and coronal images of the 8” Catphan, 
MTF module with bead source #CTP445. 

 

Detail Algorithm 

MTF (z-axis) is 12.4 lp/cm @ 10%, 15 
lp/cm @ 0% which is a 0.333 mm limiting 
resolution 

Contiguous Helical Coverage 

In Plane Helical Scan Image Quality 

Multiple helical scans can be performed in 
succession with only 5-second delays 
between helical scans, providing up to 
2,881 contiguous rotations (with up to 
1,500 images in one series). 

For details of scan technique please refer 
to the technical reference manual. 

High-Contrast Spatial Resolution 

On GE Performance phantom: 

(The following chart assumes 120-second 
helical with 5-sec IGD) 

Standard Algorithm 

* Option 

Summary of Contents for LightSpeed 16

Page 1: ...o provide an optimized workflow solution from acquisition to final report Lowest dose acquisition using SmartmA TM an automatic modulation technique and DLP display that tells the operator the dose before the scan starts HyperPlane TM reconstruction algorithm tackles the inherent issues with fast high pitch helical scanning by reducing helical artifacts and dose and optimizing the slice profile Cr...

Page 2: ... may allow you to reduce contrast media perform better thin slice CT angiography exams use thinner slices for most exams and perform longer helical exams without tube cooling delays Remote Gantry Tilt from the Operator s Console to increase exam speed including built in safety features to prevent accidental contact of the gantry with the patient CTDIw DLP Dose Length Product and Dose Efficiency di...

Page 3: ...exam to the patient All console components monitors keyboard mouse and BrightBox are free standing and can be easily moved on the front work surface Large 1024 color display Patient demographics and exam protocols can be preprogrammed in advance of patient arrival through the Schedule Patient feature Leading edge real time image processing MPR MPVR Volume Viewer All surfaces made of high impact re...

Page 4: ...ontinue with additional helical scans or a set of axial scans Helical 8 SLICE HELICAL MODES Table Speed mm rotation Slice Thick ness mm Pitch 0 625 1 0 875 1 1 35 1 1 675 1 1 25 6 25 8 75 13 5 16 75 2 5 6 25 12 5 8 75 17 5 13 5 27 0 16 75 33 5 3 75 6 25 12 5 8 75 17 5 13 5 27 0 16 75 33 5 5 6 25 12 5 8 75 17 5 13 5 27 0 16 75 33 5 7 5 6 25 12 5 8 75 17 5 13 5 27 0 16 75 33 5 10 6 25 12 5 8 75 17 5...

Page 5: ...16 slice recon mode Ten user defined regions Each region has one default protocol displayed with the anatomical selector for very fast access to most commonly used protocols Scan Speeds Full 360 rotational scans in 0 5 0 6 0 7 0 8 0 9 1 0 seconds Maximum image to image cycle time is 10 for prospective and retrospective image to image display This applies for 512 matrix any display FOV in AutoView ...

Page 6: ...413 Prospective Recon Operator may initiate full recons at any table location in increments of 1 10 the image thickness image thickness remains constant 0 5mm Retrospective Recon Operator may initiate full recons at any table location in 0 1 mm increments image thickness remains constant Retrospective Image Decomposition The operator has the option to retrospectively decompose the original raw dat...

Page 7: ...e increment Gantry display in 0 5 mm increments On 8 inch 20 cm CATPHAN phantom 5 mm 3 at 13 3 mGy 3 mm 3 at 37 2 mGy For each rotation of the gantry the LightSpeed 16 collects 16 rows of scan data There are five reconstruction modes available for creating images from the multi slice scan data 1i 2i 4i 8i and 16i By using 1i 2i 4i and 8i reconstruction modes scan data can be combined prior to imag...

Page 8: ...r to acquisition The operator can select different reconstruction algorithms and display fields of view for each reconstruction This frees the operator from sitting at the console and directly contributes to increased productivity Standard Scans On GE Performance phantom Scan Time ISD mA Scans Acquisition Time 1 sec 1 sec 440 18 45 0 35 1 27 1 1 400 24 55 0 47 1 49 1 1 360 32 68 1 03 2 15 1 1 340 ...

Page 9: ...aker microphone and volume controls CTDIvol expressed in mGy 100 mAs Three button mouse with mouse pad The ImageWorks desktop also provides a gateway for DICOM 3 0 image transactions either through a local area network or via DICOM formatted MOD media Head 18 4 mGy 100 mAs Body 9 3 mGy 100 mAs BrightBox trackball assembly Two wide work surfaces Scout Scans All these devices are free standing and c...

Page 10: ...removed from the prescription Scan recon control uses only 2 screens to set up first scan New Patient and Protocol View Edit Dose Computation Display CTDIvol CTDI volume DLP Dose Length Product and Dose Efficiency computation and display during scan prescription provides patient dose information to the operator Protocol View Edit A single full screen View Edit table allows fast and easy examinatio...

Page 11: ...lows full use of the image display monitor for AutoView and image review processing without interruption during AutoFilm Gray Scale Enhancement An AutoFilm Composer provides a simple programming interface for automated filming set up ProView visualization algorithms are available to enhance anatomical structures without additional reconstruction time Batch Filming is accomplished through a single ...

Page 12: ...is is the ultimate tool to assess and quantify vascular structures including stenosis analysis stent planning procedures post stenting or vascular surgery follow up protocol driven tools to perform quick flexible and accurate quantitative analysis of vascular anatomy provides maximum minimum and mean intraluminal diameter measurements provides cross sectional areas of true orthogonal sections of t...

Page 13: ... sec and 40 mm sec System Components Horizontal Range 170 cm Horizontal Scannable Range 170 cm metal free axial and 160 cm metal free helical Scout Gantry Image transfer time using either AdvantageNet or DICOM 3 0 protocols is approximately 0 1 second per 512 image on a 100baseT Network Horizontal Speed Up to 100 mm sec Advanced slip ring design continuously rotates generator tube detector and dat...

Page 14: ... disk drive Average time to replace tube 10 hours High Performance Drive 2GB DDR266 Dual Channel Memory with a throughput of 4 2GB sec 36 GB 3 5 inch form factor High Voltage Generation UltraSCSI interface Image Processor High frequency on board generator Continuous operation during scan Standard MOD drive Nvidia Quadro4 980XGL AGP 8X graphics with 128MB Memory Magnetic Optical Disk Drive 53 2 kW ...

Page 15: ...time and receive pull query and storage commitment push Siting Requirements Minimum Suite Size AdvantageNet GenesisNet point to point send receive and pull query no broadcast Overall Suite Compatible UPS options are available if desired 15 5 feet 4 74 m wide x 19 5 feet 5 96 m deep x 9 feet 2 75 m high InSite point to point Cooling Requirements TCP IP for system administration Recommended Suite Si...

Page 16: ...ight 1 887 74 1 Table Width 610 24 Length 5 050 199 Height 1 120 44 Power Distribution Unit Width 762 30 Depth 585 23 Height 1 270 50 Operator Console Width 1 238 48 7 Depth 1 288 50 7 Height 1 257 49 5 Warranty The published Company warranty in effect on the date of shipment shall apply The Company reserves the right to make changes All specifications are subject to change Regulatory Compliance T...

Page 17: ...4th 2003 Page 17 LightSpeed16 With Xtream CT Scanner System GE Medical Systems America Milwaukee USA Fax 1 414 544 3384 GE Medical Systems Asia Tokyo Japan Fax 81 425 85 5490 Hong Kong Fax 852 2559 3588 GE Medical Systems Europe Option ...

Page 18: ...htSpeed16 With Xtream CT Scanner System GE Medical Systems America Milwaukee USA Fax 1 414 544 3384 GE Medical Systems Asia Tokyo Japan Fax 81 425 85 5490 Hong Kong Fax 852 2559 3588 GE Medical Systems Europe Minimum Room Layout Typical Room Layout Option ...