GE LightSpeed 16 Product Data Sheet Download Page 5

g

 

Product Data Sheet – rev 4.2 
July 4th 2003  
Page 5 
 
 

 

LightSpeed

16

 

 

With Xtream 

CT Scanner System 

GE Medical Systems-America 
Milwaukee, USA  - Fax: 1 414 544 3384 
GE Medical Systems-Asia: 
Tokyo, Japan - Fax: 81 425 85 5490 
Hong Kong - Fax: 852 2559 3588 
GE Medical Systems-Europe 

Scan Fields of View: 

Biopsy

:  Simplified prescription for single 

or multiple scans around an arbitrary table 
position aids biopsy studies.  

 

—  25 cm for adult head 

16-SLICE HELICAL MODES 

Table Speed (mm/rotation) 

Slice 

Thick-

ness 

(mm) 

Pitch 

0.5625  

:1 

  

 

0.9375 

:1 

  

 

1.375

:1 

  

 

1. 75:1 

  

0.625 5.625 9.375 

13.75  17.5 

1.25  

5.625 

 11.25 

9.375 
18.75  

13.75  

27.5 

17.5  

35 

2.5  

5.625 

 11.25 

 9.375 

18.75  

13.75  

27.5 

17.5 

35  

3.75  

 5.625 
 11.25 

9.375 
18.75 

13.75  

27.5 

17.5 

35 

5 5.625 

 11.25 

9.375 
18.75 

13.75  

27.5 

17.5 

35 

7.5 11.25 18.75 27.5 

35 

10 11.25 18.75 27.5 

35 

—  25, 50 cm for body 

Advanced Artifact Reduction (AAR)

 

Filter significantly reduces streaking 
artifacts when highly absorbent objects 
are in the field of view – ie: large 
shoulders, screws. 

—  25 cm for pediatric head 

Helical Scan Enhancements 

Helical Image Reconstruction 

Full simultaneity allows complete image 
display, processing and analysis, as well 
as image archival and filming, concurrent 
with scanning and reconstruction -- even 
when acquiring helical images in a multi-
slice mode. 

Reconstruction Algorithms: Soft Tissue, 
Standard, Detail, Bone, Bone Plus, Lung 
and Edge 

Reconstruction Matrix: 512 

Confirm Rx to X-Rays on: < 15 sec. for 
any state of tube and gantry; < 10 sec. 
with the gantry rotating 

Display Matrix: 1024. 

Display FOV: Freely variable center/off-
center, prospective/retrospective target 
selection. 

Anatomical Programmer: a ten-region 
anatomical selector allows quick and easy 
access to 90 user-programmable 
protocols per region.  Separate selector 
for adult and pediatric exams.  There are 
four selection tabs to select: GE, User, 
Service and Most Recent Patient.  
Copy/Paste is supported for easy 
modification and copying of protocols. 

CT Number Scale: -1024 to 3071 HU 

Helical Reconstruction Times: 

 

 

Reconstruction time as fast 6 

images per second 

Helical Scan Parameters 

 

Typical 0.167 sec. image-to-image 

recon in normal 16 slice recon mode. 

Ten user-defined regions.  Each region 
has one default protocol displayed with the 
anatomical selector for very fast access to 
most commonly used protocols 

Scan Speeds: Full 360° rotational scans in 
0.5, 0.6, 0.7, 0.8, 0.9, 1.0 seconds.  

 

Maximum image-to-image cycle 

time is ± 10% for prospective and 
retrospective image-to-image display. 
This applies for 512 matrix; any 
display FOV; in AutoView (all 
layouts); with concurrent filming and 
image archival for all scan modes. 

Scan Technique: 

Protocols include preset scan time, kVp, 
mA, scan mode, slice thickness and 
spacing, table speed, scan FOV, display 
FOV and center, recon algorithm and 
special image acquisition and processing 
options   

—  kVp:   80, 100, 120, 140 

—  mA: 10 to 440 in 5 mA increments 

—  Power:  0.8 to 53.2 kW 

 

Iterative bone processing 

increases time by 250 milliseconds. 
Iterative bone processing, which is 
always enabled for adult head 
scanning, reduces image artifacts in 
head scans stemming from X-ray 
beam hardening effects. 

—  Focal Spot Selection: 

Any scan parameters may be edited for 
each scan or all scans - either before or 
during an exam.  The number of scans 
may also be easily changed. 

 

  Small spot for up to 24 kW 

 

  Large spot for greater than 24 kW 

Single Acquisition: 120 second scan 
maximum. 

AutoScan™:

 Fully automates longitudinal 

table movement and start of each scan. 

Minimum DFOV: 9.6 cm 

Helical Tilt: helical acquisition is possible 
with the gantry tilted to a maximum of 30 
degree, in half degree increments. 

AutoVoice™:

 3 preset (English) and 17 

user-defined messages automatically 
deliver patient breathing instructions; 
especially useful for multiple helical 
scanning. 

Minimum Pixel Size: 0.19 mm 

Queued Recon: Requests will be 
processed continuously and 
simultaneously with other processes on 
the system including scanning.  
Prospective recon will be prioritized over 
retrospective recon. 

Multiple Acquisition Maximum Scan Time: 
Multiple scans may be acquired in one 
series to produce up to 3,000 contiguous 
helical images. Up to 2,881 seconds 
helical coverage is possible in multiple 
series. 

Trauma Patient

: Allows patient scans and 

image display/analysis without entering 
patient data before scanning. 

Priority Recon Queuing: One touch 
selection marks most recent rotation for 

Minimum Inter-Group Delay (IGD): 5 
seconds between adjacent helical scans 

* Option 

Summary of Contents for LightSpeed 16

Page 1: ...o provide an optimized workflow solution from acquisition to final report Lowest dose acquisition using SmartmA TM an automatic modulation technique and DLP display that tells the operator the dose before the scan starts HyperPlane TM reconstruction algorithm tackles the inherent issues with fast high pitch helical scanning by reducing helical artifacts and dose and optimizing the slice profile Cr...

Page 2: ... may allow you to reduce contrast media perform better thin slice CT angiography exams use thinner slices for most exams and perform longer helical exams without tube cooling delays Remote Gantry Tilt from the Operator s Console to increase exam speed including built in safety features to prevent accidental contact of the gantry with the patient CTDIw DLP Dose Length Product and Dose Efficiency di...

Page 3: ...exam to the patient All console components monitors keyboard mouse and BrightBox are free standing and can be easily moved on the front work surface Large 1024 color display Patient demographics and exam protocols can be preprogrammed in advance of patient arrival through the Schedule Patient feature Leading edge real time image processing MPR MPVR Volume Viewer All surfaces made of high impact re...

Page 4: ...ontinue with additional helical scans or a set of axial scans Helical 8 SLICE HELICAL MODES Table Speed mm rotation Slice Thick ness mm Pitch 0 625 1 0 875 1 1 35 1 1 675 1 1 25 6 25 8 75 13 5 16 75 2 5 6 25 12 5 8 75 17 5 13 5 27 0 16 75 33 5 3 75 6 25 12 5 8 75 17 5 13 5 27 0 16 75 33 5 5 6 25 12 5 8 75 17 5 13 5 27 0 16 75 33 5 7 5 6 25 12 5 8 75 17 5 13 5 27 0 16 75 33 5 10 6 25 12 5 8 75 17 5...

Page 5: ...16 slice recon mode Ten user defined regions Each region has one default protocol displayed with the anatomical selector for very fast access to most commonly used protocols Scan Speeds Full 360 rotational scans in 0 5 0 6 0 7 0 8 0 9 1 0 seconds Maximum image to image cycle time is 10 for prospective and retrospective image to image display This applies for 512 matrix any display FOV in AutoView ...

Page 6: ...413 Prospective Recon Operator may initiate full recons at any table location in increments of 1 10 the image thickness image thickness remains constant 0 5mm Retrospective Recon Operator may initiate full recons at any table location in 0 1 mm increments image thickness remains constant Retrospective Image Decomposition The operator has the option to retrospectively decompose the original raw dat...

Page 7: ...e increment Gantry display in 0 5 mm increments On 8 inch 20 cm CATPHAN phantom 5 mm 3 at 13 3 mGy 3 mm 3 at 37 2 mGy For each rotation of the gantry the LightSpeed 16 collects 16 rows of scan data There are five reconstruction modes available for creating images from the multi slice scan data 1i 2i 4i 8i and 16i By using 1i 2i 4i and 8i reconstruction modes scan data can be combined prior to imag...

Page 8: ...r to acquisition The operator can select different reconstruction algorithms and display fields of view for each reconstruction This frees the operator from sitting at the console and directly contributes to increased productivity Standard Scans On GE Performance phantom Scan Time ISD mA Scans Acquisition Time 1 sec 1 sec 440 18 45 0 35 1 27 1 1 400 24 55 0 47 1 49 1 1 360 32 68 1 03 2 15 1 1 340 ...

Page 9: ...aker microphone and volume controls CTDIvol expressed in mGy 100 mAs Three button mouse with mouse pad The ImageWorks desktop also provides a gateway for DICOM 3 0 image transactions either through a local area network or via DICOM formatted MOD media Head 18 4 mGy 100 mAs Body 9 3 mGy 100 mAs BrightBox trackball assembly Two wide work surfaces Scout Scans All these devices are free standing and c...

Page 10: ...removed from the prescription Scan recon control uses only 2 screens to set up first scan New Patient and Protocol View Edit Dose Computation Display CTDIvol CTDI volume DLP Dose Length Product and Dose Efficiency computation and display during scan prescription provides patient dose information to the operator Protocol View Edit A single full screen View Edit table allows fast and easy examinatio...

Page 11: ...lows full use of the image display monitor for AutoView and image review processing without interruption during AutoFilm Gray Scale Enhancement An AutoFilm Composer provides a simple programming interface for automated filming set up ProView visualization algorithms are available to enhance anatomical structures without additional reconstruction time Batch Filming is accomplished through a single ...

Page 12: ...is is the ultimate tool to assess and quantify vascular structures including stenosis analysis stent planning procedures post stenting or vascular surgery follow up protocol driven tools to perform quick flexible and accurate quantitative analysis of vascular anatomy provides maximum minimum and mean intraluminal diameter measurements provides cross sectional areas of true orthogonal sections of t...

Page 13: ... sec and 40 mm sec System Components Horizontal Range 170 cm Horizontal Scannable Range 170 cm metal free axial and 160 cm metal free helical Scout Gantry Image transfer time using either AdvantageNet or DICOM 3 0 protocols is approximately 0 1 second per 512 image on a 100baseT Network Horizontal Speed Up to 100 mm sec Advanced slip ring design continuously rotates generator tube detector and dat...

Page 14: ... disk drive Average time to replace tube 10 hours High Performance Drive 2GB DDR266 Dual Channel Memory with a throughput of 4 2GB sec 36 GB 3 5 inch form factor High Voltage Generation UltraSCSI interface Image Processor High frequency on board generator Continuous operation during scan Standard MOD drive Nvidia Quadro4 980XGL AGP 8X graphics with 128MB Memory Magnetic Optical Disk Drive 53 2 kW ...

Page 15: ...time and receive pull query and storage commitment push Siting Requirements Minimum Suite Size AdvantageNet GenesisNet point to point send receive and pull query no broadcast Overall Suite Compatible UPS options are available if desired 15 5 feet 4 74 m wide x 19 5 feet 5 96 m deep x 9 feet 2 75 m high InSite point to point Cooling Requirements TCP IP for system administration Recommended Suite Si...

Page 16: ...ight 1 887 74 1 Table Width 610 24 Length 5 050 199 Height 1 120 44 Power Distribution Unit Width 762 30 Depth 585 23 Height 1 270 50 Operator Console Width 1 238 48 7 Depth 1 288 50 7 Height 1 257 49 5 Warranty The published Company warranty in effect on the date of shipment shall apply The Company reserves the right to make changes All specifications are subject to change Regulatory Compliance T...

Page 17: ...4th 2003 Page 17 LightSpeed16 With Xtream CT Scanner System GE Medical Systems America Milwaukee USA Fax 1 414 544 3384 GE Medical Systems Asia Tokyo Japan Fax 81 425 85 5490 Hong Kong Fax 852 2559 3588 GE Medical Systems Europe Option ...

Page 18: ...htSpeed16 With Xtream CT Scanner System GE Medical Systems America Milwaukee USA Fax 1 414 544 3384 GE Medical Systems Asia Tokyo Japan Fax 81 425 85 5490 Hong Kong Fax 852 2559 3588 GE Medical Systems Europe Minimum Room Layout Typical Room Layout Option ...

Reviews: