Product Data Sheet – rev 4.2 
July 4th 2003  
Page 14 
 
 

 

LightSpeed

16

 

 

With Xtream 

CT Scanner System 

GE Medical Systems-America 
Milwaukee, USA  - Fax: 1 414 544 3384 
GE Medical Systems-Asia: 
Tokyo, Japan - Fax: 81 425 85 5490 
Hong Kong - Fax: 852 2559 3588 
GE Medical Systems-Europe 

Peripherals: 

• 

Heat Storage Capacity: 6.3 MHU 

• 

16 x 1.25 mm (uses all 24 rows) 

• 

16 x 0.625 mm (uses center 16 rows )  

• 

Heat Dissipation:   

Total of 254 GB system: 

• 

Anode (max) 840 KHU/min 

70% geometric efficiency; 98% absorption 
efficiency. 

• 

Casing (cont) 300 KHU/min 

• 

Main system (host) disk drive: 

• 

Tube Unit: 6.9 kW continuous for 10 
minutes 

• 

High Performance Drive 

Data Acquisition System 

• 

36 GB, 3.5 inch form factor 

• 

Dual Focal Spots: 

• 

15,000 RPM 

• 

Small Focal Spot: 

12,288 available input channels. 

• 

Ultra320 SCSI interface 

0.7  mm (W) x 0.6 mm (L) nominal 
value  
(IEC 336/93)

 

• 

Assigned to applications software and 
image files 

1,640 Hz maximum sample rate. 

Effective analog to digital conversion range 
greater than two million to one. 

• 

2 system disk drives (Image Disk) 

0.9 mm (W) x 0.7 mm (L) traditional 
methodology

 

• 

High Performance Drive 

• 

73 GB, 3.5 inch form factor each. 

Scan/Control Unit 

• 

15,000 RPM 

• 

Large Focal Spot: 
0.9 mm (W) x 0.9 mm (L) nominal 
value 
(IEC 336/93)

 

• 

Ultra320 SCSI interface 

Located in base of Operator Console. 

• 

Assigned to image files only 

 

- 250,000 uncompressed 512 images  

1.2 mm (W) x 1.2 mm (L) 
traditional methodology 

• 

Scan data disk drive: 

Host Computer 

• 

High Performance Drive 

• 

36 GB, 3.5 inch form factor 

• 

Maximum Power: 53.2 kW 

• 

Dual SMP 2.66.GHz Intel Xeon processors 
with 512KB L2 cache. 

• 

UltraSCSI interface  

• 

Beam collimated to 55° fan angle. 

• 

Intel Hyper-threading technology. 

• 

Scan data disk drive: 

Average time to replace tube: < 10 hours 

• 

High Performance Drive 

• 

2GB DDR266 Dual Channel Memory with a 
throughput of 4.2GB/sec 

• 

36 GB, 3.5 inch form factor 

High Voltage Generation 

• 

UltraSCSI interface 

Image Processor: 

 

• 

High-frequency on-board generator. 
Continuous operation during scan.  

• 

Standard 

MOD

 drive: 

• 

Nvidia Quadro4 980XGL AGP 8X graphics 
with 128MB Memory 

• 

Magnetic Optical Disk Drive 

• 

53.2 kW output power. 

• 

Erasable, rewritable media 

• 

kVp: 80, 100, 120, 140 

• 

Graphics Processor Unit (GPU) Clock 
300Mhz 

• 

2.3 GB, 3.5 inch form factor 

• 

mA: 10 to 440 mA, 5 mA increments 

• 

Assigned to DICOM 3.0 format image 
file. 

• 

Graphics Memory Clock 325Mhz

 

Maximum mA for each kVp selection: 

 

• 

Stores 4,700 lossless JPEG compressed 
512 image files per side 

 

Image Reconstruction Engine 
(GRE) 

KVp Max 

mA 

80 400 

100 420 
120 440 
140 380 

• 

Off-line retrieval of image. Images may 
be viewed as soon as they are restored 
from MOD 

 

 

• 

Custom-designed special purpose 
CT Image Generator 

• 

DVD Ram

• 

Custom CT back projection hardware 
accelerates 2D & 3D back projection. 

• 

4.7 GB per side, 5.25" half height form 
factor 

HiLight Matrix II Detector: 

• 

Intel Hyper-threading Technology.

 

• 

Transfer rate 2.7MB/sec 

• 

32-bit floating point data format

 

• 

Assigned to scan data file and protocol 
file storage/retrieval. 

21,888 individual elements composed by: 8 
rows of 1.25 mm thickness and 16 rows of 
0.625 mm thickness, each containing 888 
active patient elements; 24 reference 
elements. 

• 

2GB DDR226 ECC Dual Channel 
Memory Standard (4.2 GB/sec).

 

 

• 

Color monitors

Software Architecture: 

• 

21 inch diagonal width 

4 typical modes of data output: 

• 

1280 x 1024 dot resolution 

• 

Software architecture based on industry 
standards and client-server design 

• 

8 x 1.25 mm (uses center 16 rows) 

• 

Non-interlaced, flicker-free presentation 

• 

8 x 2.5 mm (uses all 24 rows) 

• 

76 kHz Horizontal deflection frequency 

* Option 

Summary of Contents for LightSpeed 16

Page 1: ...o provide an optimized workflow solution from acquisition to final report Lowest dose acquisition using SmartmA TM an automatic modulation technique and DLP display that tells the operator the dose before the scan starts HyperPlane TM reconstruction algorithm tackles the inherent issues with fast high pitch helical scanning by reducing helical artifacts and dose and optimizing the slice profile Cr...

Page 2: ... may allow you to reduce contrast media perform better thin slice CT angiography exams use thinner slices for most exams and perform longer helical exams without tube cooling delays Remote Gantry Tilt from the Operator s Console to increase exam speed including built in safety features to prevent accidental contact of the gantry with the patient CTDIw DLP Dose Length Product and Dose Efficiency di...

Page 3: ...exam to the patient All console components monitors keyboard mouse and BrightBox are free standing and can be easily moved on the front work surface Large 1024 color display Patient demographics and exam protocols can be preprogrammed in advance of patient arrival through the Schedule Patient feature Leading edge real time image processing MPR MPVR Volume Viewer All surfaces made of high impact re...

Page 4: ...ontinue with additional helical scans or a set of axial scans Helical 8 SLICE HELICAL MODES Table Speed mm rotation Slice Thick ness mm Pitch 0 625 1 0 875 1 1 35 1 1 675 1 1 25 6 25 8 75 13 5 16 75 2 5 6 25 12 5 8 75 17 5 13 5 27 0 16 75 33 5 3 75 6 25 12 5 8 75 17 5 13 5 27 0 16 75 33 5 5 6 25 12 5 8 75 17 5 13 5 27 0 16 75 33 5 7 5 6 25 12 5 8 75 17 5 13 5 27 0 16 75 33 5 10 6 25 12 5 8 75 17 5...

Page 5: ...16 slice recon mode Ten user defined regions Each region has one default protocol displayed with the anatomical selector for very fast access to most commonly used protocols Scan Speeds Full 360 rotational scans in 0 5 0 6 0 7 0 8 0 9 1 0 seconds Maximum image to image cycle time is 10 for prospective and retrospective image to image display This applies for 512 matrix any display FOV in AutoView ...

Page 6: ...413 Prospective Recon Operator may initiate full recons at any table location in increments of 1 10 the image thickness image thickness remains constant 0 5mm Retrospective Recon Operator may initiate full recons at any table location in 0 1 mm increments image thickness remains constant Retrospective Image Decomposition The operator has the option to retrospectively decompose the original raw dat...

Page 7: ...e increment Gantry display in 0 5 mm increments On 8 inch 20 cm CATPHAN phantom 5 mm 3 at 13 3 mGy 3 mm 3 at 37 2 mGy For each rotation of the gantry the LightSpeed 16 collects 16 rows of scan data There are five reconstruction modes available for creating images from the multi slice scan data 1i 2i 4i 8i and 16i By using 1i 2i 4i and 8i reconstruction modes scan data can be combined prior to imag...

Page 8: ...r to acquisition The operator can select different reconstruction algorithms and display fields of view for each reconstruction This frees the operator from sitting at the console and directly contributes to increased productivity Standard Scans On GE Performance phantom Scan Time ISD mA Scans Acquisition Time 1 sec 1 sec 440 18 45 0 35 1 27 1 1 400 24 55 0 47 1 49 1 1 360 32 68 1 03 2 15 1 1 340 ...

Page 9: ...aker microphone and volume controls CTDIvol expressed in mGy 100 mAs Three button mouse with mouse pad The ImageWorks desktop also provides a gateway for DICOM 3 0 image transactions either through a local area network or via DICOM formatted MOD media Head 18 4 mGy 100 mAs Body 9 3 mGy 100 mAs BrightBox trackball assembly Two wide work surfaces Scout Scans All these devices are free standing and c...

Page 10: ...removed from the prescription Scan recon control uses only 2 screens to set up first scan New Patient and Protocol View Edit Dose Computation Display CTDIvol CTDI volume DLP Dose Length Product and Dose Efficiency computation and display during scan prescription provides patient dose information to the operator Protocol View Edit A single full screen View Edit table allows fast and easy examinatio...

Page 11: ...lows full use of the image display monitor for AutoView and image review processing without interruption during AutoFilm Gray Scale Enhancement An AutoFilm Composer provides a simple programming interface for automated filming set up ProView visualization algorithms are available to enhance anatomical structures without additional reconstruction time Batch Filming is accomplished through a single ...

Page 12: ...is is the ultimate tool to assess and quantify vascular structures including stenosis analysis stent planning procedures post stenting or vascular surgery follow up protocol driven tools to perform quick flexible and accurate quantitative analysis of vascular anatomy provides maximum minimum and mean intraluminal diameter measurements provides cross sectional areas of true orthogonal sections of t...

Page 13: ... sec and 40 mm sec System Components Horizontal Range 170 cm Horizontal Scannable Range 170 cm metal free axial and 160 cm metal free helical Scout Gantry Image transfer time using either AdvantageNet or DICOM 3 0 protocols is approximately 0 1 second per 512 image on a 100baseT Network Horizontal Speed Up to 100 mm sec Advanced slip ring design continuously rotates generator tube detector and dat...

Page 14: ... disk drive Average time to replace tube 10 hours High Performance Drive 2GB DDR266 Dual Channel Memory with a throughput of 4 2GB sec 36 GB 3 5 inch form factor High Voltage Generation UltraSCSI interface Image Processor High frequency on board generator Continuous operation during scan Standard MOD drive Nvidia Quadro4 980XGL AGP 8X graphics with 128MB Memory Magnetic Optical Disk Drive 53 2 kW ...

Page 15: ...time and receive pull query and storage commitment push Siting Requirements Minimum Suite Size AdvantageNet GenesisNet point to point send receive and pull query no broadcast Overall Suite Compatible UPS options are available if desired 15 5 feet 4 74 m wide x 19 5 feet 5 96 m deep x 9 feet 2 75 m high InSite point to point Cooling Requirements TCP IP for system administration Recommended Suite Si...

Page 16: ...ight 1 887 74 1 Table Width 610 24 Length 5 050 199 Height 1 120 44 Power Distribution Unit Width 762 30 Depth 585 23 Height 1 270 50 Operator Console Width 1 238 48 7 Depth 1 288 50 7 Height 1 257 49 5 Warranty The published Company warranty in effect on the date of shipment shall apply The Company reserves the right to make changes All specifications are subject to change Regulatory Compliance T...

Page 17: ...4th 2003 Page 17 LightSpeed16 With Xtream CT Scanner System GE Medical Systems America Milwaukee USA Fax 1 414 544 3384 GE Medical Systems Asia Tokyo Japan Fax 81 425 85 5490 Hong Kong Fax 852 2559 3588 GE Medical Systems Europe Option ...