Immunity Test
IEC 60601-1-2 Test Level Compliance Level
Electromagnetic Environment
Note: P is the power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the rec‐
ommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by
an electromagnetic site survey,* are less than the compliance
level in each frequency range.**
Interference may occur in the vicinity of equipment marked
with the following symbol:
NOTE:
* Field strengths from fixed transmitters, such as base stations for cellular telephones
and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast
cannot be estimated accurately. To assess the electromagnetic environment due to fixed
RF transmitters, an electromagnetic site survey should be performed. If the measured
field strength exceeds the RF compliance level above, observe the Innova system to
verify normal operation in each use location. If abnormal performance is observed,
additional measures may be necessary, such as re–orienting or relocating the Innova
system
**Over the frequency range 150 kHz to 2.5GHz, the recommended field strengths is here
less than 0.3V/m.
The Recommended Separation Distances are listed in
.
These are guidelines. Actual conditions may vary.
2.3.3.3 Recommended Separation Distances for Portable and Mobile RF Communications Equipment
IEC 60601–1–2
Table 3-10:
Frequency of Transmitter
150 KHz to 26 MHz
26 MHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
Equation
d = [3.5 / V2]√P
d = [3.5 / V2]√P
d = [3.5 / E2]√P
d = [3.5 / E4]√P
Rated Power of Trans‐
mitter (watts)
Distance (meters)
Distance (meters)
Distance (meters)
Distance (meters)
10 mW
1.2
1.2
1.2
2.3
100 mW
3.8
3.8
3.8
7.3
1
12
12
12
23 (*)
10
38
38
38
73
100
120
120
120
230
For transmitters rated at a power not listed above, the DISTANCE can be estimated using the
equation in the corresponding column, where P is the power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
NOTE:
These are guidelines. Actual conditions may vary.
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Direction 5177951-4-1EN, Revision 1
62